SR Instruments SR Scales SR241KG Operating And Service Manual Download Page 9

Model SR241KG Portable Infant/Adult Scale 
Service Manual  
Part No. MAN241KG_201006 

 

Page 9 of 14 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

         Copyright 2020

 

S

Instruments, Inc. 

BASIC SYSTEM OPERATION  

ADULT WEIGH 

STEP 1

:  Remove  infant  cradle  from  base.    Place  scale  on  hard  smooth  floor  surface  to  ensure 

accurate data.  

 

STEP 2

:  Press the large blue “

ZERO

” button and wait for the display to indicate “

0.0

”  

STEP 3

:  Position  patient  on  the  platform.    Weight  will  be  displayed  in  a  few 

seconds.   

STEP 4

:  To  recall  last  weight  data,  press  the  “

KG

”  button.   

Note

:  Must  be  done  before 

ZERO

” button is pressed. 

STEP 5

:  If more weight  needs  to  be added at  any time during the weighing process, 

add  weight  and  press  the  blue  “

ADULT  WEIGH

”  button.  This  will  update  the  net 

weight on the display 

 

BABY WEIGH – OPTION 1 

STEP 1

:  If necessary, attach the baby cradle by sliding and snapping it into place. 

STEP 2

:  Place the platform on a stable counter or table-top. 

STEP 3

:  Follow the directions for 

ZERO-BABY MODE

 in the previous section. 

STEP 4

:  When the display indicates “

0.0

”, you may place the baby on the scale, press 

the  “

KG

” button to recall the last stored weight data.  

 

BABY WEIGH – OPTION 2 

STEP 1

:  Remove  infant  cradle  from  base.    Place  scale  on  hard  smooth  floor  surface  to  ensure 

accurate data.   

STEP 2

:  Have an adult stand on the scale holding any accessories that will be necessary 

for the baby’s comfort (blanket, diaper, etc.).  Press the yellow “

BABY  ZERO

” button.  

This will zero the scale system and TARE the adult’s and additional accessories weight. 

STEP 3

:  Place  the  baby  in  the  adult’s  arms.    The  weight  displayed  will  be  the  infant’s  net 

weight.   

STEP 4

:  To  recall  last  weight  data,  press  the  “

KG

”  button.   

Note

:  Must  be  done  before 

ZERO

” button is pressed. 

STEP 5

:  If more weight needs to be added at any time during the weighing process, add 

weight and press the yellow “

BABY WEIGH

” button. This will update the net weight on 

the display. 
 
 
 
 
 

Summary of Contents for SR Scales SR241KG

Page 1: ...ge 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instruments In...

Page 2: ...ES 2 PACKING CHECKLIST MODEL SR241KG 3 PACKING CHECKLIST MODEL SR241SOKG 3 ASSEMBLY 4 REPLACEMENT PARTS AND ACCESSORIES 4 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE 5 CLEANING AND DISINFECTING 6 SP...

Page 3: ...truments Inc PACKING CHECKLIST Model SR241KG Portable Infant Adult Scale PACKING CHECKLIST Model SR241SOKG Portable Scale DESCRIPTION QUANTITY SCALE BASE 1 ea INFANT CRADLE 1 ea 9 VOLT BATTERY 1 ea CE...

Page 4: ...t Cradle slides onto the Scale Base and snaps securely into position REPLACEMENT PARTS and ACCESSORIES SYSTEM DESCRIPTION and INTENDED USE SYSTEM DESCRIPTION The SR241KG SR241SOKG Portable Infant Adul...

Page 5: ...ensures accurate and repeatable weight readings The SR241SOKG Portable Adult Scale is a unique portable scale system that is lightweight and weather resistant It weighs a mere two kilograms This scal...

Page 6: ...se a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfecta...

Page 7: ...CY 0 2 1 digit of displayed resolution over 0 2 kg AUTO ZERO One button operation SR241KG SCALE WEIGHT 3 kg 7 lb SR241SOKG SCALE WEIGHT 2 kg 4 lb AUTO POWER DOWN After 1 minute AVERAGING Automatic dig...

Page 8: ...be displayed ZERO BABY MODE To zero for infant weigh Slide and snap the cradle into place on the scale Place any blankets pads or disposable liners being used for infant s comfort on the cradle before...

Page 9: ...sary attach the baby cradle by sliding and snapping it into place STEP 2 Place the platform on a stable counter or table top STEP 3 Follow the directions for ZERO BABY MODE in the previous section STE...

Page 10: ...2020 SInstruments Inc BATTERY REPLACEMENT REPLACING BATTERIES The display will read LobAt when battery is low and needs to be changed STEP 1 Figure 3 Turn the scale over and locate the battery cover...

Page 11: ...ty voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unloaded system The output of the operational...

Page 12: ...including the caregiver is touching the patient or scale Set scale on smooth hard floor surface Scale reads 999 9 Return to factory for repair For additional information or assistance phone our Servic...

Page 13: ...rranties expressed or implied and of all other obligations or liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special dama...

Page 14: ...No MAN241KG_201006 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc...