VII. Using the iSpin-NMR System
Testing iSpin-NMR: Step-by-step Instructions
The iSpin-NMR is tested extensively before shipment and is ready to use for acquisition of NMR (or
NQR) signals out-of-the-box, however, it is suggested that the user tests the iSpin-NMR upon
delivery to make sure there were no damages during shipment and to become more familiar with
the operation and performance of individual components.
The testing procedure for the iSpin-NMR system is described below:
1. Download and install the latest SpinAPI package available at:
http://spincore.com/support/spinapi/
Additionally, very detailed information on each programming function can be found on
our SpinAPI References page that is linked from our main SpinAPI page – the direct link
http://www.spincore.com/support/spinapi/reference/production/2013-09-
On the above page, please select spinapi.h for detailed descriptions of each and every
SpinAPI function.
The RadioProcessor Examples are part of the SpinAPI package.
2. Connect your iSpin-NMR system as shown in Figure 7. For iSpin's with SMA connectors
that match Figure 5, the 'To Probe' connector is the 'RF Power Out' connector and the
left 'λ/4' connector is the 'NMR Signal In' connector. Note that pin-4 on the TTL output
DB-9 connector is the blanking signal and can be used to trigger the oscilloscope.
Turn on your iSpin-NMR system only after making all connections.