Indoor Air Quality Meter 

 

 

800046 

 

Instruction Manual 

Summary of Contents for 800046

Page 1: ...Indoor Air Quality Meter 800046 Instruction Manual ...

Page 2: ...TS Introduction 3 Features 4 Power Supply 5 LCD Display 6 Keypad 7 Measurement Procedures 8 12 CO2 Calibration 13 RH Calibration 14 15 Troubleshooting 16 PC Connection 17 CO2 Levels and Guidelines 18 Specifications 19 ...

Page 3: ...est air from HVAC equipment Simultaneously backlit display of CO2 carbon dioxide level humidity and air temperature or the user may choose a rotating display of dew point and wet bulb temperature instead of air temperature Also calculates TWA Time Weighted Average and STEL Short Term Exposure Limit Simple user calibration of CO2 and RH Features audible C02 threshold alarms min max ave hold and a U...

Page 4: ...ble CO2 warning alarm Battery and adapter power supply Easy manual calibration of CO2 and humidity PC connect via USB serial interface Statistics of weighted averages TWA 8 hours weighted average STEL 15 minutes weighted average Meter 4 AA Alkaline Batteries Instruction Manual Hard Carrying Case MATERIALS SUPPLIED www ponpe com sale ponpe com ...

Page 5: ...teries are firmly seated 3 Re install the cover and ensure it snaps securely into place Note When the AC Adapter is used power supplied from the batteries is disabled The adapter does not charge the batteries When the battery voltage is low Lob appears on LCD and an alarm sounds Under low battery conditions the meter s CO2 sensor will not function properly readings are not displayed and an audible...

Page 6: ...ighted average 8 hours Short term exposure limit 15 minutes weighted average Readings are frozen Minimum Maximum readings Low battery indicator Dew point temperature Air temperature Wet bulb temperature Relative humidity Celsius Fahrenheit Measurement icons www ponpe com sale ponpe com ...

Page 7: ...tup Page mode Enters CO2 calibration with Enters RH calibration with Freezes current readings Turns LCD Backlight On and Off Selects unit or increases Setup value Selects AIR DP or WBT temperature displays Selects unit or decreases Setup value Activates MIN MAX STEL and TWA functions Saves settings www ponpe com sale ponpe com ...

Page 8: ... its readings every second In environments where conditions change i e from low to high temperature it takes the meter s sensor 30 seconds to respond for CO2 and 30 minutes for RH Note Do not hold the meter close to your face while taking measurements because exhaling affects CO2 levels Power On Off Press DP WBT to switch temperature modes The lower left of the LCD display cycles from air temperat...

Page 9: ...unction and resume measurement Backlight Minimum Maximum Short Term Exposure Limit and Time Weighted Average 1 While in Normal mode press MN MX AV to view min imum maximum and weighted average readings 2 Press MN MX AV to display MIN MAX STEL and TWA readings in sequence and then return to Normal mode 3 In MIN and MAX modes the unit displays minimum and maximum CO2 readings on the upper LCD and th...

Page 10: ...ter is turned on for a period shorter than 15 minutes STEL value readings will indicate the weighted average of readings sampled since the unit was turned on This rule also applies to TWA when the meter is on for a period shorter than 8 hours It takes at least 5 minutes to calculate STEL and TWA readings During the first 5 minutes the LCD displays While all readings are frozen STEL and TWA are upd...

Page 11: ...ing SET then press and hold SET and HOLD simultaneously for 2 seconds until n appears on the LCD Note Auto Power Off is disabled when the unit is in the Cal ibration mode To enter Setup mode ensure the meter is in Normal mode and press and hold SET for more than 1 second To exit Setup mode press CAL with either P 1 0 or P 3 0 displayed and the meter will return to Normal mode 1 When entering Setup...

Page 12: ... is not applicable for the 800046 but has been reserved for future models 1 While in P1 0 press MODE or DP WBT to access P 3 0 and choose the temperature scale 2 Press MN MX AV and the meter will go into the P 3 1 mode with its current setting for C or F displayed 3 To switch between C and F scales press MODE or DP WBT 4 Press MN MX AV to save your setting or press CAL to exit without saving and r...

Page 13: ...ne exhaust Procedure 1 Take the meter to an appropriate calibration area 2 Turn the meter on and simultaneously hold down CAL and MODE to enter CO2 calibration mode 3 400ppm and CAL will blink on the LCD while cal ibration is being performed 4 In approximately 10 minutes blinking stops calibration is completed and the meter returns to Normal mode 5 To abort calibration turn the meter off at any ti...

Page 14: ...ge Contact Sper Scientific for calibration salt or services RH CALIBRATION 33 Calibration 1 Plug the sensor probe into a 33 salt bottle 2 With the meter in Normal mode hold CAL and DP WBT simultaneously to enter the 33 calibration mode 3 CAL and the calibrating value 32 7 if at 25 C will blink on the LCD with the current temperature displayed on the left 4 Meter calibration is completed in approxi...

Page 15: ...t 25 C blink on the LCD 4 The calibration process will be completed in approxi mately 60 minutes when CAL and humidity stop blinking Note Users can also calibrate individually either to 33 or 75 To calibrate 33 only press CAL and exit when 33 cal ibration is competed To calibrate 75 only press MODE or DP WBT within the first 5 minutes of initializing the 33 calibration process www ponpe com sale p...

Page 16: ... the Data Hold function Slow Response Ensure that the air flow channels located at the top rear of the unit are not blocked E01 CO2 sensor is damaged E02 The value is under range E03 The value is over range E04 The temperature or humidity data error results in this DP WB error E07 Voltage too low to measure CO2 Replace batteries or use adapter E11 Retry humidity calibration E17 Retry CO2 calibrati...

Page 17: ...ng and data analysis via USB serial interface and software The protocol is 9600 bps 8 data bits no parity Format ASCII Cxxxxppm Txxx xC F Hxx x dxxx xC F wxxx xC F LRC CRLF Description CO2 Air RH DP WBT LRC CRLF PC CONNECTION www ponpe com sale ponpe com ...

Page 18: ...Institutes of Health 5 000ppm or 30 000ppm 15 min ACGIH 5 000ppm or 30 000ppm 15 min The American Conference of Government Industrial Hygienists ACGIH Note TWA Time Weighted Average stands for the average carbon dioxide level exposure during 8 hours working day is 5 000ppm 5 days STEL Short Term Exposure Limit value shows the last 15 minutes CO2 concentration is 30 000ppm ASHARE Standard 62 1989 S...

Page 19: ... 0 1 Accuracy 3 10 90 5 others Dew Point Range 13 3 60 0 C 8 140 F Resolution 0 1 C 0 1 F Wet Bulb Range 73 4 60 0 C 100 140 F Resolution 0 1 C 0 1 F Warm Up 30 seconds Operating 0 50 C 0 95 RH avoid condensation Storage 20 60 C 0 99 RH avoid condensation Power 4pcs AA Alkaline Batteries AC Adapter Battery Life 24 hours Alkaline Dimensions Weight 8 x 3 x 2 210 x 75 x 65mm 10oz 280g Optional Access...

Page 20: ...eries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will void the warranty To obtain warranty service ship the unit postage prepaid to SPER SCIENTIFIC LTD 7720 E Redfield Rd Suite 7 Scottsdale AZ 85260 WWW SPERSCIENTIFIC COM INFO SPERSCIENTIFIC COM The defective unit must be accompanied by a description of the...