7  Input and Output Connections 

 

7.1 

Monitoring and Control 

7.1.1  HV enable 

This is a digital input which controls the HV output.  Driving this input low (< 0.8V) 
will cause the HV output to be enabled and ramp up to its programmed value. The 
filament current will start to ramp up after the high voltage has reached its 
programmed value, approximately 3 seconds after the HV output is enabled. 
Disconnecting or driving this input high (>2.4V ) will cause the filament to be turned 
off and the HV output disabled. The maximum input voltage of this input is 12V. 

7.1.2  Filament stable 

Open collector digital output indicating that the filament current has stabilised and 
the tube could be producing X-rays. This should occur 8 

 10s after the HV enable is 

activated. A low output (<0.8V) indicates that the filament current has not stabilised 
The maximum rating of this input is 12V.  

7.1.3  HV voltage program output 

A 10V internal 

reference voltage and 12 turn 20KΩ potentiometer is provided that 

can be used to locally control or pre-set the high voltage output. Connecting this 
output to the high voltage program input will allow the high voltage output to be pre 
set/controlled using the internal potentiometer. 

7.1.4  HV voltage program input 

 10V input corresponding to 0 

 10kV output. Ac/-2% of full scale. Z

in

 = 

5MΩ. 

 

7.1.5  HV voltage monitor 

 10V output corresponding to 0 

 10kV output. Ac/-2% of full scale. Z

out

 = 

10kΩ.

 

7.1.6  Emission current monitor 

 10V output corresponding to 0 

 1mA emission current. Ac/-3% of full 

scale. Z

out

 

= 10kΩ.

 

7.1.7    Filament current monitor 

 10V output corresponding to 0 

 5A filament current. Ac/-5% of full 

scale. Z

out

 

= 2k2Ω.

 

7.1.8  Filament current Pre heat (set value) 

Provided for adjustment of the filament Pre heat current level. 0 

 5V output 

corresponding to a Pre heat current setting of 0.5A 

 2.5A filament current. 

 

7.1.9  Filament current Preset Maximum filament current (set value) 

  Provided for adjustment of the filament current limit value. 0 

 10V output 

corresponding to a maximum filament current of 0 

 5A. 

7.1.10  Emission current program input 

 10V input corresponding to 0 

 1mA emission current. Ac/-3% of full 

scale. Z

in

 

= 1MΩ.

 

7.1.11  Emission current program output 

A 10V internal reference voltage and 12 turn 5KΩ potentiometer is provided that can 

be used to control or pre-set the emission current. Connecting this output to the 
emission current program input will allow the emission current output to be pre-set 
using the internal potentiometer. 

7.1.12  Pin connections 

The monitoring and control is provided through a 15 way ‘D’ connector which also 

provides input power to the unit. 

 
 

 

Summary of Contents for XMPG10P10

Page 1: ...ION INSTRUCTIONS Document Number 81233 4 Issue A 1 Date 23 01 18 16 04 2021 Issuing Authority I 5650 10465C Checked Approved Tel 44 0 1798 873986 Fax 44 0 1798 872479 E mail hvsales spellmanhv co uk Broomers Park Broomers Hill Lane Pulborough W Sussex RH20 2RY ...

Page 2: ...quipment is turned off and the capacitance of both the load and power supply are grounded Allow adequate time for discharge of internal capacitance of the power supply Do not ground yourself or work under wet or damp conditions Servicing Safety Maintenance may require removing the Instrument cover with the power on Servicing should only be done by qualified personnel aware of the hazards If in dou...

Page 3: ...fications Added environmental conditions Added mechanical Added input and output connections details Added setup and adjustments 1 Contents 1 Unit Description 4 2 Safety 4 3 Regulatory Specifications 4 4 Environmental conditions 5 5 Mechanical 5 6 Installation 6 7 Input and Output Connections 7 8 Set Up Adjustments 9 ...

Page 4: ...for use in a Pollution Degree 2 Installation Category II environment Consideration should be given to conducting the following tests with the unit installed in the end product Dielectric Voltage Withstand Test between live parts of the unit and the end product chassis Permissible Limits Tests with the unit installed in the end product Temperatures on power electronic components transformer winding...

Page 5: ...te humidity maximum 25 gm Maximum storage period 0 25 year at above conditions Long term storage 3 year special packaging required 5 Mechanical 5 1 Mechanical outline 185mm x 105mm x 33 5mm The mass of the module is nominally 2kg 105 0 33 5 185 0 66 5 9 7 30 0 23 7 46 0 14 7 15w D plug HOLE Ø4 5mm IN 3 POSITIONS MOUNTING HOLE TEMPLATE VIEWED FROM THE TOP OF UNIT NOTE Max thread depth into unit is ...

Page 6: ...gree 2 Installation Category II environment The input and output connectors are not intended for field connections and should only be connected to internal wiring in the end product The unit is intended for use as a component and no surface of the unit should be accessible in the end product 6 3 Electrical Installation The unit must be terminated safely before operation Hazardous voltages will be ...

Page 7: ...r 7 1 4 HV voltage program input 0 10V input corresponding to 0 10kV output Accuracy 2 of full scale Zin 5MΩ 7 1 5 HV voltage monitor 0 10V output corresponding to 0 10kV output Accuracy 2 of full scale Zout 10kΩ 7 1 6 Emission current monitor 0 10V output corresponding to 0 1mA emission current Accuracy 3 of full scale Zout 10kΩ 7 1 7 Filament current monitor 0 10V output corresponding to 0 5A fi...

Page 8: ... 0 10V output 11 Filament current monitor 0 10V output 12 Emission current program input 0 10V input 13 Emission current program output 0 10V from internal pre set 14 Signal ground 15 Pre set maximum filament current set value 0 10V from internal pre set The filament Preheat level and current limit are set by internal presets accessible through the case side If external high voltage control is not...

Page 9: ... heat current 0 5V corresponds to 0 5A 2 5A filament current 8 6 If external HV control is required connect the control signal to pin 8 with return to pin 14 0 10V corresponds to 0 10kV 8 7 If internal HV control is required link pins 7 and 8 then turn the HV preset adjustment on the case front whilst monitoring the voltage at pin 8 return to pin 14 Set for the HV potential required 8 8 If externa...

Reviews: