LPX-200
Industrial Imaging System
3-3
Chapter 3: Installing the LPX-200 X-ray System
System Interconnections - Air Cooled
The following details the connections for setting up a liquid
cooled LPX-200 X-ray System.
! WARNING !
All cables MUST be connected to their
appropriate connectors on the
Control Unit, Cooling Unit, and
Tubehead before applying power to
the system.
1.
Install the Cooling Fan power cable:
◆
Connect the male end of the power cable to the
connector labeled
“
Cooler
”
on the Control Unit.
◆
Connect the female end of the power cable to the
connector on the Tubehead base plate.
2.
Install the Tubehead Control cable:
◆
Connect the male end of the Control cable to the
connector labeled
“
Tubehead
”
on the Control Unit.
◆
Connect the female end of the Control cable to the
connector on the Tubehead base plate.
3.
Install the line power cable:
◆
Connect the female end of the line power cable to the
connector labeled
“
Power
”
on the Control Unit.
◆
Connect the
“
plug
”
end of the line power cable to the
AC voltage source (see Connecting to Power
instructions later in this section).
4.
Make the External Interlock connection:
◆
If available, connect the Interlock cable to the
connector labeled
“
Interlock
”
on the Control Unit
(see External Interlock instructions later in this
section).
◆
Connect the
“
jumper
”
(supplied) to the connector
labeled
“
Interlock
”
on the Control Unit for units that
do not employ an external interlock system.
Figure 3-2:
LPX-200 Set Up (Air Cooled)
Line Power Cable
to AC Source
CONTROL
UNIT
TUBEHEAD
Cable
8 pin connector
Cable
10 pin connector