pH 400, pH 600 Meter (item 2107, 2108)
Quick Start Guide
View full manual at
www.specmeters.com
Keypad (p. 8)
•
The meter is turned on and off with the power bu on (1).
•
Navigate through the meter screens with the arrow bu ons (2).
•
Select, confirm, or reject op'ons with the Enter (3) and Back (4) bu ons.
Connect probe (p. 7)
•
Align the white notches on the probe connector and meter socket, having the
notch facing towards the front side of the meter.
•
The connector is pushed down un'l it clicks into the meter’s socket.
Note:
Do not turn the probe connector once inserted into the meter’s socket.
When removing the probe, pull on the connector, not the cable.
Ba#eries (p. 6)
•
The ba ery compartment is accessed by removing the screw on the back of the
meter. The meter is shipped with 3 fully charged , rechargeable AA ba eries. To
ac'vate the meter for the first 'me, pull the plas'c strip (ba ery isolator) out of the
ba ery compartment and press the power bu on.
•
To recharge the ba eries, connect the meter to the power adapter using the
supplied USB cable and place the power adaptor in a power outlet. The meter will
recharge in approximately 7 hours.
•
It is possible to use non-rechargeable ba eries.
Warning!
Do not connect the meter to the power adaptor or to a USB when non-rechargeable, alkaline
ba eries are installed. This may damage the ba eries and/or meter.
For
pH 600, select the Spectrum buffer set (p. 24)
•
Select the
Buffer Selec on
op'on from the
Calibrate
menu.
•
Select
Sentron buffers
.
•
Press the Home bu on
NOTE:
pH 400 meter defaults to the Sentron buffer set.
Sentron buffers
SEN
SEN