![Spectris SERVOPRO Chroma Operator'S Manual Download Page 32](http://html.mh-extra.com/html/spectris/servopro-chroma/servopro-chroma_operators-manual_1344053032.webp)
SERVOPRO CHROMA Gas Analyser
31
04400001A / Revision 7
3
SPECIFICATIONS
WARNING
You must install and use the SERVOPRO Chroma in accordance
with the requirements of this section and subsequent sections of the
manual. If you do not, the protection facilities incorporated into the
design of the analyser may not operate as intended, sample gas
measurements may not be accurate, or the analyser may be damaged.
WAR
CAUTION
To comply with the applicable requirements of EU EMC Directive: The case of the
Auxiliary oven shall be bonded to the case of the Master/Secondary unit using braid less
than 20 cm long.
WAR
CAUTION
To comply with the applicable requirements of EU EMC Directive: The case of the
Standalone PC shall be bonded to the case of the Master/Secondary unit using braid less
than 20 cm long.
WAR
CAUTION
To comply with the applicable requirements of EU EMC Directive: The assembly shall be
bonded to a local EMC earth. The bond shall be less than 1 m in length.
CAUTION
The Master, Secondary, Standalone PC and Auxiliary oven enclosures do not contain any
user replaceable parts inside.
CAUTION
If the analyser is subjected to external RF signals; additional errors can occur:
FID: up to 5% full scale
Plasma: up to 3% full scale
TCD: up to 5% full scale