SERVOPRO CHROMA Gas Analyser
113
04400001A / Revision 7
7.4 Shut-down procedure
1.
Press quit button in software and enter analyser password. Wait for screen to
turn black and turn-off power.
CAUTION
It is important to allow the analyser to power itself down. Turning off the
power to the analyser after the screen turns black but before it completes
the shutdown procedure may result in loss of data.
2.
Wait for one hour to let column cool down with all gases connected.
3.
Disconnect all gases.
WARNING
Do not install caps on vent if carrier or sample gas inlets are still
connected. This will damage the detector.
4.
Install caps on all inlets and outlets.