2-14
MiniCapt® Mobile Operations Manual
Chapter 2
: Unpacking and Installation
Installation
Adjust the Plate Height in the Plate Holder
To ensure the MiniCapt Mobile operates at its optimal performance, adjust the plate
holder to ensure the agar plate is in the proper placement.
The MiniCapt Mobile can be used with any agar plate having the following dimensions:
•
A
diameter
from 86 to 92mm
•
An
agar volume
from 18 to 32ml
•
An
agar depth
from 3 to 6mm
Plate height adjustments
WARNING
Remove power to the unit prior to removing sampler head.
>> To adjust for the plate height:
1.
Remove the sampler head by turning counterclockwise.
WARNING
After removing the sampling head, take care to ensure loose items
are not dropped into the opening to the blower.
Figure 2-13
Sampler head on
MiniCapt
Figure 2-14
Sampler head off
of the MiniCapt
Figure 2-15
Sampler head
removed