18-6
MiniCapt® Mobile Operations Manual
Chapter 18
: Accessories
Remote Connection to BioCapt
Remote Connection to BioCapt Single-Use
The Remote Connection to BioCapt can be used to attach the MiniCapt Mobile
Microbial Air Sampler to a Single-Use Impactor.
The Remote Connection to BioCapt Single-Use may only be used with 25/50 LPM units.
To install the Remote Connection to BioCapt Single-Use Kit on the MiniCapt:
1.
Follow steps 1 and 2 of
on page 4-2 to install the agar plate.
Remove the sampler head and expose the plate holder (see
2.
Place the
Remote Sampling Adaptor
on the MiniCapt.
3.
Place the
Single-Use Impactor
inside the isolator or at the suitable sampling
point.
4.
Connect the
Impactor
to the
Remote Sampling Adaptor
:
a.
Use a Tri-Clamp to attach one end of the tubing to the
Remote Sampling
Adaptor
.
b.
Attach the other end of the tubing to the
Single-Use Impactor
.
Figure 18-5
BioCapt Single-Use
NOTE:
BioCapt Single-Use can be purchased separately. Contact your Particle
Measuring Systems representative for more information.