![Spectris PARTICLE MEASURING SYSTEMS MiniCapt Operation Manual Download Page 112](http://html.mh-extra.com/html/spectris/particle-measuring-systems-minicapt/particle-measuring-systems-minicapt_operation-manual_1344052112.webp)
Audio for Warnings and Alarms
Chapter 8
: Warnings and Alarms Setup
MiniCapt® Mobile Operations Manual
8-13
7.
The external alarm activation is driven using the compressed gas auxiliary port.
The auxiliary port powers any external device working at a 12VDC with a
maximum current of 10W (800mA).The external alarm connection requires the
connector product number (PN) 1000019268.
+12VDC = pin 2
GND = pin 3
Figure 8-23
Connection diagram for auxiliary external alarm connection