Function Codes
Available public functions are:
•
3
: Read Holding Registers
•
4
: Read Input Registers
•
16
: Write Multiple Registers
Please note that function codes 3 and 4 are fully interchangeable as they
behave in the same way.
Data Formats
Float
The Float refers to the floating point according to IEEE 754 (Single Precision).
This format requires two registers obtaining 32 bit where each register contains
the high byte in its first bit.
For example, if the float value is 20.56 (int32), represented as 0x41A47AE1
(hexaint32), it is written in two consecutive registers, where the first register
is 3499. Therefore, the value has to be transmitted in the following way:
Int32
All int32 values are 32bit-wide integer values. The example given in the
previous section applies here as well.
Character
The definition is as follows:
•
8bit ASCII-Code table according to ISO-8859-1 (Latin-1 Western
European)
Function code 16 may be used with broadcast (device ID = 0).
Codes 3 and 4 may not be used with broadcast.
Register 3499, high byte
0x7A
Register 3499, low byte
0xE1
Register 3500, high byte
0x41
Register 3500, low byte
0xA4
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...