Operation
OXY5500 Operator’s Manual
3
–
37
•
“Error code: 1” = No RTD (Pt100) (bit O)
•
“Error code: 5” = No RTD (Pt100) and amplitude too low (bit 0 [2
N
Value 1], bit 2 [2
N
Value 4)=5)
•
“Error code: 1024” = No pressure sensor connected (bit 10)
•
“Error code: 1029” = No RTD (Pt100), amplitude too low, no pressure
sensor connected (bit 0 [2
N
Value 1], bit 2 [2
N
Value 4], bit 10 [2
N
Value 1024] = 1029)
Table 3–2
Error Codes
Bit
2
N
Value
Error
0
1
No RTD (Pt100)
1
2
No sensor selected
2
4
Amplitude too low
3
8
SD card defect
4
16
Reference amplitude out of range
5
32
Photo diode saturated
6
64
Signal overflow
7
128
Signal overflow
8
256
Reserved
9
512
Critical error, contact
10
1024
No pressure sensor / pressure sensor out of range
11
2048
Reserved
12
4096
Storage space full
Summary of Contents for OXY5500
Page 1: ...P N 4900002239 rev G OXY5500 Oxygen Analyzer Operator s Manual ...
Page 2: ......
Page 8: ...Oxygen Analyzer iv 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 12: ...Oxygen Analyzer viii 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 22: ...Oxygen Analyzer 1 10 4900002239 rev G 7 11 17 THIS PAGE IS INTENTIONALLY LEFT BLANK ...
Page 34: ...Oxygen Analyzer 2 12 4900002239 rev G 3 4 16 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 86: ...Oxygen Analyzer 3 52 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 108: ...Oxygen Analyzer 4 22 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...
Page 116: ...Oxygen Analyzer A 8 4900002239 rev G 7 11 17 THIS PAGE INTENTIONALLY LEFT BLANK ...