
©
Spectradyne LLC, All rights reserved
35
Version 2.5.0 Revision 3, July 2020
3.
Combine and Compare
At this point in the workflow, the Currently Loaded file list typically contains multiple Stats files
that together represent a polished analysis of the particle content of a single physical sample. The
goals of this workflow step are to:
1.
Combine the multiple Stats files into a single entity for simpler manipulation.
2.
Use the Viewer’s quantitative analysis tools to compare the measurement to others.
3.1
Combine multiple Stats files into a single Combined file
Combine Procedure
➢
Load the desired Stats files into the Currently Loaded file list.
➢
Navigate to the Data Output tab of the Spectradyne Viewer.
➢
In the Combine Data and Generate Run Report
frame, use the “All Loaded” button to
:
o
Combine all loaded Stats files into a single Combined file.
o
Generate a 1-page Run Report (Excel format) summarizing the measurement. See
page 36 for more details about the Run Report.
➢
Two new files will be generated in the location specified by the user in the export dialog
window:
o
A Combined file: F
ile name of the form, “
<prefix>_
YYYYMMDD_ccNN.h5”, where
NN is an incrementing number unique for each day.
o
A Run Report: Same file name as Combined file, with .xlsx extension.
3.2
Compare Multiple Analyses
➢
Load the analyses of interest into the Currently Loaded file list.
➢
In the QuickView Processed Data frame of the Spectradyne Viewer, select Plot Mode Multi.
➢
Use the tools in the Data Visualization and Quantification tabs of the Viewer to make
quantitative comparisons of the analyses.
➢
Scatter and CSD plots will display separate results from each loaded analysis on the same
axes, and quantitation tools will report results for each analysis separately.
Quantifying Particle Size Distributions
–
Concentration in the CSD
➢
The concentration of particles over a certain size range is readily obtained after plotting
the CSD (Figure 19):
❖
In the Quantification tab of the Viewer, select the size range of interest either
graphically or by manually entering the minimum and maximum values
❖
“Graphical”
–
Click Select and drag the mouse cursor across the CSD over the size
range of interest.
❖
“Manual”
–
Enter the desired min and max sizes and click Select
❖
Click “
Concentration
” to
report the concentration over the selected size range.
❖
The concentration measurement is displayed on the plot.