Finish Priming Manually
This automatic sequence interrupts the automated cartridge priming routine and allows the user
to finish priming the cartridge using non-standard pressures.
➢
Use only when the cartridge is not priming normally because of low surfactant content in
the sample, or very high viscosity
➢
This is an advanced user feature and special training is required
—
please contact
Spectradyne’s support team for more details (
Run Cleaning Cartridge Only
This automatic sequence flushes the internal fluid lines of the nCS1 and requires a reusable
cleaning cartridge.
➢
Click the
“
Run Cleaning Cartridge
”
button and follow the prompts.
➢
Wait for the routine to complete.
➢
Remove the reusable cleaning cartridge and store it in a safe place for future use.
Drain Bottle Lines
This automatic sequence drains any remaining liquid in the lines back into the bottles
➢
Click the
“
Drain Bottle Lines
”
button and follow the prompts.
➢
Wait for the routine to complete.
Check Supply Pressure
This automatic sequence is used to ensure sufficient pressure is supplied to the instrument.
➢
DO NOT USE WHEN MEASUREMENTS ARE BEING PERFORMED
➢
Click “Check Supply Pressure” and wait for the dialog wind
ow to appear
➢
The measured pressure should be > 10 PSI
Restarting the Auto-analysis Engine
In the rare event where the Auto-analysis engine freezes or fails to process the incoming data and
the Live CSD stops updating during acquisition,
click the “Restart Auto
-
analysis Engine” button
to
reboot the engine.
About this instrument
…
Use this button to obtain information about the connected instrument. A dialog window will be
displayed with Hardware Version Number and Serial Number. If the Hardware Version Number is
0, then Serial Number defaults to 0.