Instrument Shutdown
This automatic sequence flushes the internal fluid lines of the nCS1 and subsequently prepares
the instrument for shutdown. It requires a reusable cleaning cartridge and deionized water.
➢
P
ress the “
Instrument Shutdown
” button
(Figure 9) to begin the automated process.
➢
Follow the prompts:
o
Load a reusable cleaning cartridge
o
Click OK to prepare for shutdown
➢
Wait for first stage of process to complete. Follow the prompt to replace running buffer
with DI water (bottles 1 and 3):
o
Disconnect the running buffer bottles from the fluid lines using the quick-connect
tube fittings. Refrigerate running buffer if intended for future use.
o
Add 25-50 mL of deionized water to running buffer bottles or separate DI water
bottles and replace in the instrument.
Note: Use only fresh deionized water
o
Click OK to continue the shutdown routine
➢
Follow the next prompt to remove the cleaning cartridge and store it in a safe place for
re-use.
o
Click OK to complete the routine.
➢
IMPORTANT:
o
Refrigerate buffer bottles and DI water bottles after use
Less/More
The “More” button opens an extra panel on the right for
diagnostic/troubleshooting purposes.
➢
To hide the Diagnostic
panel, press the “Less”
button
Diagnostic Panel
➢
This panel is used for diagnostic purpose and troubleshooting measurement issues. It can
be revealed
by clicking the “More” button under the “Utilities” frame
, Figure 9. Please
contact Spectradyne at