![Spectracom GSG-5 series User Manual With Scpi Manual Download Page 233](http://html1.mh-extra.com/html/spectracom/gsg-5-series/gsg-5-series_user-manual-with-scpi-manual_1343630233.webp)
6.3
Command Reference
6.3.1
Common Commands
6.3.1.1
*CLS
Clear Status Command
The
*CLS
common command clears the status data structures by clearing all event registers and
the error queue. Also possible executing of scenario or signal generator is stopped. It does not
clear enable registers and transition filters. It clears any pending *WAI, *OPC, and *OPC?.
Example
*CLS
6.3.1.2
*ESE
Standard Event Status Enable
Sets the enable bits of the standard event enable register. This enable register contains a mask
value for the bits to be enabled in the standard event status register. A bit that is set true in the
enable register enables the corresponding bit in the status register. An enabled bit will set the
ESB (Event Status Bit) in the Status Byte Register if the enabled event occurs.
Command Syntax
*ESE <decimal>
Parameters
<dec.data> = the sum (between 0 and 255) of all bits that are true.
Event Status Enable Register (1 = enable)
Bit
Weight
Enables
7
128
PON, Power-on occurred
6
64
URQ, User Request
5
32
CME, Command Error
4
16
EXE, Execution Error
3
8
DDE, Device Dependent Error
6.3 Command Reference
CHAPTER
6
•
User Manual GSG-5/6 Series
Rev.
26
217
Summary of Contents for GSG-5 series
Page 2: ......
Page 4: ...Blank page II User Manual GSG 5 6 Series...
Page 116: ...BLANK PAGE 3 6 Options Menu 100 CHAPTER 3 User Manual GSG 5 6 Series Rev 26...
Page 182: ...BLANK PAGE 4 9 Studioview Tasks 166 CHAPTER 4 User Manual GSG 5 6 Series Rev 26...
Page 368: ...BLANK PAGE 6 7 Revision History SCPI Guide 352 CHAPTER 6 User Manual GSG 5 6 Series Rev 26...