6. Operation
6.1 Connections
Using a Cat5 STP cross-over cable, or standard patch cable with the supplied adaptor fitted, connect
the host PC’s network interface card to the Cirrus using the “RGA Comms” connector.
Connect the “Power” socket to the mains supply using a suitable power lead.
6.2 Powering up
Switch on the Cirrus.
The internal cooling fans will start, but none of the pumps should be running, the only indicators lit on
the front panel should be “Power” and a blinking “Comms” Led. This is due to interlocking preventing
the powering up of components until the turbo pump is switched on and is up to speed.
Use the RGA Device Manager to locate and install your unit ready for use. This is explained in the
Process Eye Professional User Manual.
Once the unit has been installed, start the Process Eye software, once loaded you will see the unit
information is displayed.