CONTROL UNIT SETTINGS
1. Turn unit ON at the switch provided on the lower left-hand side of the air-control unit.
2. Select the desired
THERAPY MODE
.
- For Standard and XL models, select
FLOAT
or
ALTERNATING
as desired.
- For LR model, select
FLOAT
or
ALTERNATING
or
ROTATION
.
3.
AIR FLOW
- Select OFF or ON as desired. Air flow ON is recommended where prevention of skin maceration
is desired.
4. LOCKOUT –
This feature can be used to eliminate accidental or
unintended changes to the control unit settings. To engage, press
LOCKOUT firmly and hold for three to four seconds until indicator light
turns on. To disengage, press LOCKOUT again and hold firmly for 3-4
seconds until light turns off, indicating that the control unit settings are
again unlocked and can be adjusted as desired.
5.
COMFORT LEVEL:
For best possible pressure management, maximize immersion and envelopment by
initially
setting the “COMFORT LEVEL” on the control panel to the softest selection appropriate for the
patient, in accordance with the stated weight limits:
Easy Air, Easy Air LR:
•
For patients up to 120 lbs., begin with level 1 (“Max. Immerse”).
•
For patients weighing more than 120 lbs. (up to 500 lbs.) begin with Level 2.
Adjust for user comfort as desired, using up and down arrow buttons.
Easy Air XL:
•
Because it is not necessary for bariatric patient, the XL model does not include
“Max. Immerse” setting. Begin in lowest setting and adjust for comfort.
NOTE
: Allow the mattress to inflate for about 20 minutes before putting a patient on it.