Use only the following recommended cleaning solutions:
•
Mild soap and water solution
•
U.S. Pharmacopoeia (USP) green soap
•
Sodium hypochlorite solution (1:10 dilution of household chlorine
bleach in water)
•
Phenolic germicidal detergent (1% aqueous solution)
•
Glutaraldehyde (2.4%) (Cidex)
•
Isopropyl alcohol (70% solution)
Note:
Over time, repeated use of a chlorine bleach solution may cause
some colors to fade.
Tape adhesive can be removed with Spacelabs Healthcare adhesive
tape remover pads (P/N 392196-001).
Questions and concerns about cleaning issues should be directed to
a Spacelabs Healthcare field service engineer.
Physical Inspection
Check the following areas for wear and repair as necessary.
Transmitter
•
Battery contacts
•
Electrode wire sockets
•
Case door
•
Case
•
NIBP adapter cable receptacle (90343/91343 only)
•
SpO
2
adapter cable receptacle (90343/91343 only)
•
Battery compartment SIP switchzes (if present)
Receivers
•
Pull handle
•
Antenna connectors
•
Communications port (SDLC connector on module rear)
Housing
•
Antenna connectors
•
Module connectors