Optional Sample Extractor Assembly
FTS Systems LyoStar™ II
40
Rev 012, 04/11
© SP Scientific 2011
Operation
Note: Make sure that the front plate is closed and the inert gas/vacuum break valve is closed.
Activate the auxiliary vacuum pump to decrease the pressure in the
1.
antechamber.
After the gauge stabilizes at the maximum position, allow the vacuum pump to
2.
operate for an additional 15 minutes.
Open the antechamber isolation door. With locking handle in the up, or
3.
unlocked position, the antechamber door can be slid open.
When antechamber isolation door is fully open, lower the handle to lock the
4.
isolation door in position.
Manipulate the selector arm to select and extract the vial(s) of choice. Vials
5.
are securely held in spring loaded jaws. Push the button to open the jaws and
release to close them.
Place the sample(s) in the antechamber directly beneath the stoppering ram.
6.
Raise handle up to
unlock the door
Push to
open the jaws
Remove shroud
to stopper vials
Lower handle to
lock shut
Press to stopper
Summary of Contents for LYOSTAR II
Page 1: ...LYOPHILIZER OPERATOR S MANUAL FTS SYSTEMS LYOSTAR II STANDARD...
Page 2: ......
Page 6: ...iv Rev 012 04 11 SP Scientific 2011...
Page 29: ...FTS Systems LyoStar II Operating Your Lyophilizer Rev 012 04 11 21 SP Scientific 2011...
Page 38: ...Data Management FTS Systems LyoStar II 30 Rev 012 04 11 SP Scientific 2011...
Page 52: ...Optional Isolation Valve FTS Systems LyoStar II 44 Rev 012 04 11 SP Scientific 2011...
Page 58: ...Optional Cleanroom Configuration FTS Systems LyoStar II 50 Rev 012 04 11 SP Scientific 2011...
Page 66: ...General Maintenance FTS Systems LyoStar II 58 Rev 012 04 11 SP Scientific 2011...
Page 70: ...Appendix A Troubleshooting FTS Systems LyoStar II 62 Rev 012 04 11 SP Scientific 2011...
Page 71: ......