FTS Systems FlexiCool
Operation
Rev 001, 10/13
11
© SP Scientific 2013
Control
When the system is running, use the Up and Down buttons (
) to
1.
bring up the setpoint.
Press the button SET ( ) to accept the setting.
2.
Powering Off
Toggle the main power switch to the OFF position.
1.
The refrigeration will shut down and the display will be blank.
2.
Unplug the unit.
3.
Thoroughly clean and dry the probe head after each use.
4.
Delivery Line
The FTS Systems FlexiCool has a delivery line that carries the refrigerant to the probe
and back. This flexible delivery line is made of corrugated stainless steel tubing with
several layers of insulation. Although this line is strong, it can be damaged to the point
of rupture if it is repeatedly bent too sharply or forcibly twisted. This line must remain
gas-tight in order to contain the refrigerant charge in the unit.
Never bend the delivery line more sharply than a 5-inch (12.7 cm) radius or a
10-inch (25.4 cm) diameter.
Supporting the delivery line at one or more places is recommended to prevent
sagging or stressing the line.
Wide contact points on the supports are recommended to avoid compressing
or degrading the internal insulation.
Adjusting the delivery line when it is cold may damage the insulation or
stainless steel tubing, which may result in reduced performance.
Note:
Do not move the delivery line when it is cold. Always turn the system off for at least
4 hours before making any adjustments to the placement of the delivery line.
Summary of Contents for FTS FLEXICOOL IMMERSION COOLERS FC100
Page 1: ...IMMERSION COOLERS FC50 FC100 OPERATOR S MANUAL FTS SYSTEMS FLEXICOOL IMMERSION COOLERS ...
Page 2: ......
Page 6: ...iv Rev 001 10 13 SP Scientific 2013 ...
Page 16: ...Getting Started FTS Systems FlexiCool 8 Rev 001 10 13 SP Scientific 2013 ...
Page 22: ...Operation FTS Systems FlexiCool 14 Rev 001 10 13 SP Scientific 2013 ...
Page 30: ...Maintenance FTS Systems FlexiCool 22 Rev 001 10 13 SP Scientific 2013 ...
Page 37: ......