Quick-Check-PowerSaver Manifold 6 Port South-Tek Systems
Revision: 0
Revision Date: 8/2/17
Page 16 of 30
2.
Clean dry compressed air (20.9% O2) can be used for the high point.
3.
Make sure to only flow 1-2 scfh to the sensor. Higher flow may damage the sensor.
4.
Never block the outlet purge line
5.6
ALARMS/FILTER PARAMETERS
The “Alarm Info” menu displays the O2 alarm setpoint on the right with the purity alarm, flow alarm, and filter alarm
enable/disable buttons on the left. Local audible alarm can be enabled/disabled to produce an audible beep in the
Quick-Check when enabled alarms are triggered.
Purity Alarm: The box under the “Purity Setpoint” is the user alarm setpoint. Factory default setpoint is 2.00% O2. The
“Purity Alarm” is triggered when any zone’s O2 purity reading is higher than the setpoint value.
Flow Alarm: The Flow Alarm is triggered when there is either too little or no flow coming from any or all zones.
Filter Alarm: The “Filter Alarm” is triggered when the “Date of Next Filter Change” is reached on the “Filter Info”
screen.
These alarms can be disabled by toggling the “Enable/Disable” button on the “Alarm Settings” screen. The local
audible can be “Enabled/Disabled” on the “Alarm Settings” screen as well.
Figure 7: Alarm Settings Screen