Note: when you use frequency sweep open/short correction, the dot frequency correction
function should be OFF. Please refer to 5.1 in this chapter.
5.6Multi-frequency List sweep for capacitance test
Measurement conditions:
Function: Cp-D
Level:1Vrms
The other parameters are shown in table 5-1:
Table 5-1 the related parameters of the capacitance test
frequency
Compared parameter
upper limit
lower limit
1kHz
Cp
325.0nF
333.0nF
10kHz
D
0.0001
0.0003
100kHz
D
0.0060
0.0100
Beep: HIGH
Alarm mode: out
Procedures:
1) Turn on the ST2826.
2) Set the basic parameters:
a)
Press [LCRZ], LCR MEAS DISP page will be displayed.
b)
The FUNC field is Cp-D, the LEVEL field is 1.000V. At the same time,
MEAS DISP, BIN NO, BIN COUNT, LIST SWEEP and MORE 1/3 will
be displayed in the softkey area.
c)
Press MORE 1/3. MEAS SETUP, CORRECTION, LIMIT TABLE, LIST
SETUP and MORE 2/3 will be displayed in the softkey area.
d)
Press LIST SETUP. LIST SWEEP SETUP page will be displayed.
e)
Move the cursor to (sweep parameter) field. The field displays FREQ
[HZ].
f)
Press [ ].Move the cursor to the parameter area of sweep point 1. The
parameter area displays .
g)
Press [1], +1 will be displayed in the bottom line of the screen. The
softkey area will display the usable units (Hz, kHz and MHz). Press kHz.
The value of this field will be changed to 1.0000k.
h)
Press [ ], move the cursor to the LMT field of sweep point. The field
displays . The softkey area will display LIMIT DATA A, LIMIT
DATA B and OFF.
i)
Press LIMIT DATA A to select the main compared parameter Cp, the LMT