Service Manual: FUSION EQUINE DR® |
81
4. Configuring the FUSION EQUINE DR
®
X-ray
System
Site Information
The site information is preconfigured at the factory.
Figure 36: Site Information tab
Configuring Panels
This chapter describes how to configure the flat-panel detectors that are supported for use with
this system.
Procedure
1.
Install the Ready Indicator (PN: 10-426) into the USB port on the Tablet
2.
Open the
Management
Displaying the Management screen
on page
64, for instructions.
The
Config
screen is displayed automatically.
3.
Click the
Panel Configuration
tab.