Operation
28
Ultrasonic Generator | DocNr.: 03-4655 R03.01 | 22.09.2015 | en-US
Preparing for treatment
6 Operation
Prior to initial and each subsequent treatment the ultrasonic generator must be
cleaned and disinfected (see chapter
Prior to each treatment functional testing of the ultrasonic generator must be per-
formed and the ultrasonic generator's power settings must be checked (see chapter
on page 29). Only then may treatment be started.
During treatment, keep an identical or comparable system ready for use in case the
first one fails.
Following treatment, disposable accessories must be properly disposed of. The ultra-
sonic generator, the instrument and the accessories must be reprocessed (see chap-
ter
6.1 Preparing for treatment
6.1.1 Adjusting the volume
An activation tone signals that ultrasonic power is being delivered to the instrument.
The volume of the activation tone can be set using the volume control. Adjustment of
the volume only affects the activation tone. The setting has no effect on the volume
of the alarm tones.
The activation tone can be turned off by turning the volume control all the way to the
left. It is however not recommended to turn the volume all the way down.
For functional testing the volume control has to be turned all the way to the right.
Procedure
1. To lower the volume, turn the volume control to the left.
To raise the volume, turn the volume control to the right.
– The volume has now been adjusted.