GasLab Q2 displays and operating the device
7
GasLab Q2
G a s Q u a l i t y A n a l y z e r
Page 53 of 135
Error during start-up
In extremely rare cases, an error can occur after or during start up, e.g. after recording a
parameterization and reboot
or an internal connection failure. If this happens, the “Home” display
with a one-way street sign is shown.
Figure 7.7: Display Q2 Not Ready
This sign indicates that only basic functionalities for error treatment are active; all the other
functions are deactivated.
The causes can vary, interruptions may have formed inside the device, or a wrong parameter has
caused the malfunction.
The error is often caused by a faulty parameterization. Correct the error by
transferring correct parameterization into the device, e.g. the backup to
factory settings (
For details see chapter 8)
If the problem
can’t be resolved by the parameterization, you should contact
Elster GmbH. Never open the housing!