Guidance and manufacturer’s declaration – electromagnetic immunity
The AC-120MD is intended for use in the electromagnetic environment specified below. The customer or the user of the
AC-120MD should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the AC-120MD,
including cables, than the recommended separation
distance calculated from the equation appliance to the
frequency of the transmitter.
Recommended separation distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
outside ISM bands
c
3 Vrms
d
= 1.2
√
P
6 Vrms
150 kHz to 80 MHz
in ISM bands
c
6 Vrms
Radiated RF
3 V/m
3 V/m
IEC 60601-1-2: 2007
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 2.3
√
P
800 MHz to 2.5 GHz
IEC 60601-1-2: 2014
d
= 2.0
√
P
80 MHz to 2.7 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should
be less than the compliance level in each frequency
range.
b
Interference may occur in the vicinity of equipment
marked with following symbol:
IEC 61000-4-3
80 MHz to 2.7 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Summary of Contents for AC-120MD
Page 63: ......
Page 72: ...规格说明 CS 10 ...
Page 73: ......
Page 82: ...規格 CT 10 ...
Page 83: ......
Page 123: ......
Page 143: ......
Page 153: ......
Page 163: ......
Page 173: ......
Page 203: ......
Page 233: ......
Page 243: ......
Page 263: ......
Page 273: ......
Page 283: ......
Page 293: ......
Page 303: ......
Page 313: ......