5.4 Power Supply at the Calculator
Supercal 5 can be supplied with either battery or external mains modules. These can be
retrofitted at any time.
5.4.1 Power Supply Modules
The mains module is equipped with a backup battery already installed.
5.5 Connecting Sensors
5.5.1 Pulse Input for Volume Counting
The Supercal 5 allows the connection of slow and fast flow sensors. For this purpose, two
specific filters are provided (normal or fast mode), which can be switched via the Superprog
software.
5.5.1.1 Technical Characteristics of the Pulse Input for Volume Counting
Designation
Description
Pulse Input
Terminal 10, flow meter 440 white cable
Terminal 11, flow meter 440 green cable
Power supply flow sensor
Terminal 9 (3.7V..3.0V, max 1mA) the flow
sensor 440 brown cable
Input frequency normal mode
Max. 5 Hz
Input frequency fast mode
Battery operation max. 200 Hz
Slow volume pulses
0,0001–99,999,999 l/Imp or Imp/l
Fast response volume electrical pulses
0,0001–99,999,999 l/Imp or Imp/l
R2
R1
C1
1M
10k
Microcontroller
+3V6
0V
.
Iin
Vin
Vout (9)
D1
10(50,52)
11(51,53)
Flow sensor
Fast flow meter
R2
R1
C1
1M
10k
+3V6
0V
Iin
Vin
Vout (9)
D1
10(50,52)
11(51,53)
Microcontroller
100k
(F > 200 Hz)
R
ext
Circuit Diagram Normal Mode
Circuit Diagram Fast Mode
Terminal 9 is intended to supply a flow sensor up to Vmax = 3.6 V and Imax = 20 µA. For high
-
er voltage or higher current, it’s strongly recommended to install a power supply module to
avoid reduced battery lifetime.