Chapter 6:
Troubleshooting
51
Chapter 6: Troubleshooting
Basic Troubleshooting
This chapter contains information to help you correct problems with system operation and provides instructions
on the proper care of the system, transducer, and accessories.
If you encounter difficulty with the system, use the information in this chapter to help correct the problem. If the
problem is not covered here, contact SonoSite technical support at the following numbers or addresses:
Technical Support (USA, Canada)
1-877-657-8118
Technical Support fax:
1-425-951-6700
Technical Support e-mail:
SonoSite website:
www.sonosite.com
and select Support
International Technical Support:
Contact your local representative or call (USA) +425-951-1330
European Service Center
+44-(0)1462-444-800
e-mail: [email protected]
Japan Service Center
+81-3-5304-5337
e-mail: [email protected]
Table 6.1: Troubleshooting
Symptom
Solution
System will not power on.
Check all power connections.
Perform the following sequence: remove DC input connector and battery;
wait 10 seconds; connect DC input or install battery; press the power key.
Ensure the battery is charged.
System image quality is poor.
Adjust the LCD screen to improve viewing angle.
Adjust the brightness, as necessary, to improve image quality.
Adjust the gain.
No CPD image.
Adjust the gain.
No Color image.
Adjust the gain or the scale.
No OB measurement selections.
Select the OB exam type.
Print does not work.
Set the correct printer in system setup.
Check the printer connections.
Check the printer to ensure that it is turned on and set up properly. See
the printer manufacturer’s instructions, if necessary.
DVD/VCR does not record.
Check the DVD/VCR connections.
Check the DVD/VCR to ensure that it is turned on and set up properly. See
applicable SonoSite accessory user guide and the manufacturers’
instructions, if necessary.
External monitor does not work.
Check the monitor connections.
Check the monitor to ensure that it is turned on and set up properly. See
the monitor manufacturers’ instructions, if necessary.
Summary of Contents for MicroMaxx
Page 1: ...MicroMaxx Ultrasound System Service Manual ...
Page 6: ...vi ...
Page 10: ...4 Chapter 1 Introduction ...
Page 56: ...50 Chapter 5 Cleaning and Disinfecting ...
Page 61: ...Chapter 6 Troubleshooting 55 Control Panel Figure 6 3 Control Panel Flow Diagram ...
Page 62: ...56 Chapter 6 Troubleshooting System Figure 6 4 System Flow Diagram ...
Page 63: ...Chapter 6 Troubleshooting 57 Battery Figure 6 5 Battery Flow Diagram ...
Page 72: ...66 Chapter 7 Replacement Procedures Figure 7 10 Power Supply Screws Screws 7 ...
Page 85: ...Appendix A Parts List 79 Figure A 1 Power Supply P05471 Figure A 2 Speaker Assembly P03872 ...
Page 89: ...Appendix A Parts List 83 Figure A 8 Main PCB Assembly P05472 ...
Page 92: ...86 Appendix B Service Event Report ...
Page 98: ...92 Index ...