SSI-6000/SSI-5800/SSI-5500/SSI-5500BW
Digital Color Doppler Ultrasound System
1.6
Transport and Storage Environmental Requirements
The following environmental conditions for transport and storage are within system
tolerances:
Temperature: -20
°C
∼
55
°C
Relative Humidity: 20%
∼
90 %, no condensation
Pressure: 700
∼
1060hPa
1.7
Electrical Requirements
Power Requirements:
110-120VAC
220-230VAC
Power Consumption:
330VA
Main unit voltage
Maintain a fluctuation range of less than
±
10% or the system may be
damaged.
Grounding
Before connecting the power cable, connect the attached ground pro-
tection cable to a specialized grounding device.
Note:
• Please ensure that the power requirements are satisfied. Only use power lines
that meet the system guidelines—failure to follow these procedures may result
in system damage.
• Voltage level may vary in different geographic locations. Adjust the voltage
rating on the rear panel according to your local voltage level.
• Use only the power button on the side panel to turn on or off the unit. Mains
power switch on the rear panel should be always ON unless prior to transport
or maintenance or service of the equipment.
1.8
Electrical Safety
Only trained health professionals should operate this system. This equipment com-
plies with the following standards:
•
The equipment conforms with the following regulations for electrical safety,
–
IEC 60601-1: 1988+A1:1991+A2:1995, Medical Electrical Equipment
Part 1: General Requirements For Safety, Class I, BF, continuous op-
eration
P/N: 4701-0061-01B