RADIOMETER PHOTOMETER
NDTM-10
INSTRUCTION MANUAL
PPUH
SONOPAN Sp. z o.o.
ul. Ciołkowskiego 2/2
15-950 Białystok, Poland
phone/fax: +48 85 742 36 62
http://www.sonopan.com.pl
July 2017
Page 1: ...RADIOMETER PHOTOMETER NDTM 10 INSTRUCTION MANUAL PPUH SONOPAN Sp z o o ul Cio kowskiego 2 2 15 950 Bia ystok Poland phone fax 48 85 742 36 62 http www sonopan com pl July 2017...
Page 2: ...TION 5 2 1 Powering the instrument 5 2 2 Battery indicator 5 2 3 Measure 6 2 4 Menu 6 2 4 1 Zeroing 7 2 4 2 Memory 7 2 4 3 Calibration 7 2 4 4 Setup 7 2 4 5 Measuring ranges 8 2 5 Turning off the inst...
Page 3: ...Part 1 General principles EN ISO 3059 Non destructive testing Penetrant testing and magnetic particle testing Viewing conditions 1 1 Equipment Basic equipment two channel measuring probe two channel i...
Page 4: ...f1 2 3 1 2 337 385nm max 365nm S 313nm 5 S 405nm 0 5 Direction matching matching quality cosine f2 1 3 cosine f2 2 3 Nonlinearity f3 1 0 5 f3 2 0 5 Measuring range 3 100 W m2 10 mW cm2 10 000 lx Reso...
Page 5: ...compartment according to the polarity marked on the back panel of the meter High capacity alkaline batteries should be used i e LR03 type The meter turns on by pushing and hold for a while button Aft...
Page 6: ...ator is displayed in front of measured value Overload is signaled when at least one actual value of the one second average is higher than measurement range In that case average value can be much less...
Page 7: ...selected On the next screen current measured value is displayed This number should be edited to obtain proper measured value using buttons to change digit value or unit prefix and button to select act...
Page 8: ...3 in section 1 4 Technical data 2 5 Turning off the instrument To turn off the instrument push and hold button for at least 2 seconds 2 6 Calibration procedure Calibration is the process of finding a...
Page 9: ...on 5 of EN ISO 9934 1 standard 5 RECOMENDATIONS FOR ISTRUMENT USE It is not recommended to expose instrument on falls shakes neither on any other factors that might cause mechanical damages Protect in...
Page 10: ...Electromagnetic compatibility The conformance to above mentioned requirements is confirmed by CE mark This product cannot be thrown away with household waste Deposit the product in an authorized elec...