PAT-806-IT
– USER MANUAL
32
Notes:
- Tested device must be turned on.
- Test circuit is electrically isolated from the mains and from PE mains lead.
- The measurement result should be read after displayed values are stabilized.
- After the measurement the tested object is automatically discharged.
4.5.1 R
ISO
measurement on devices of Class I
Connect the mains plug of the tested device into the test socket
of the meter. The measurement is made between shorted L, N
(L and L for IT) and PE. In addition, it is possible to carry out
the measurement with the probe connected to
R
ISO
-
socket.
With continuous measurement set (time =
) press and hold
START
. To block
the measurement press
ENTER
.
At the time with pre-set value it is not nec-
essary.
Signalling the presence of high-voltage
View of the screen during
measurement.
Time remaining to the end of
the measurement