DIT-10MED
– USER MANUAL
33
11 Storage
In the case of storage of the device, the following recommenda-
tions must be observed:
Make sure the meter and its accessories are dry.
In the case the meter is to be stored for a prolonged period of
time, the batteries must be removed from the device.
12 Dismantling and utilization
Worn-out electric and electronic equipment should be gathered
selectively, i.e. it must not be placed with waste of another kind.
Worn-out electronic equipment should be sent to a collection
point in accordance with the law of worn-out electric and electronic
equipment.
Before the equipment is sent to a collection point, do not dis-
mantle any elements.
Observe the local regulations concerning disposal of packages,
worn-out batteries and accumulators.
13 Technical data
IR temperature range
Temperature
measurement mode
Range
Resolution
Body
32.0…42.5ºC / 90…108ºF *
0.1ºC
0.1ºF
Surface
0…60ºC / 32…140ºF **
* Accuracy of body temperature measurement according to ASTM Standard
E1965-1998 (2003)
Range
Accuracy
32.0…35.9ºC
93.2…96.6ºF
±0.3ºC
±0.5ºF
36.0…39.0ºC
96.8…102.2ºF
±0.2ºC
±0.4ºF
39.0…42.5ºC
102.2…108.5ºF
±0.3ºC
±0.5ºF
** Accuracy of surface temperature measurement: ±0.3ºC / ±0.5ºF