DIT-10MED
– USER MANUAL
31
8
Troubleshooting
Problem
: The measured body temperature is less than
32ºC
(89.
6ºF).
Cause
: If you are in the surface measurement mode, the reading
shows the external temperature that a body generates.
Problem
: Message
HI
is displayed.
Cause
: Measurement range exceeded. Measured temperature is
higher than the upper measurement range:
42.
2ºC
(
108ºF
) for
Body
mode or
60ºC
(
140ºF
) for
Surface
mode.
Problem
: Message
LO
is displayed.
Cause
: Measurement range exceeded. Measured temperature is
lower than the lower measurement range:
32ºC
(
90ºF
) for
Body
mode or
0ºC
(
32ºF
) for
Surface
mode.
Cause of HI and LO messages
Solution
Temperature reading hampered by
hair, perspiration…
Remove factors hindering the meas-
urement.
Temperature hampered by an air flux.
Prevent the air flux.
The measuring distance is too far.
Respect the measuring distance of
5…15 cm (2…5.9”).
The instrument was moved from too
high / low temperature.
Wait for 10 minutes before taking the
measurement.
Problem
: The instrument cannot be turned on.
Solution
: Check if the batteries are correctly installed and charged.
Problem
: The display does not light up.
Solution
: Check the batteries. Replace the batteries with new ones.
Problem
: Unreadable display.
Solution
: Check the batteries. If they are depleted, replace them
with new ones.