background image

Column Oven PRSO-V1-BR

for Bruker Captive Spray

Users Manual

Revision 2016-04-14

Summary of Contents for PRSO-V1

Page 1: ...Column Oven PRSO V1 BR for Bruker Captive Spray Users Manual Revision 2016 04 14 ...

Page 2: ...2 PRSO V1 BR for Bruker Captive Spray Users Manual Sonation GmbH ...

Page 3: ...ctions 4 Scope of supply 5 Functional description 6 Oven Controller Installation 8 Hardware installation Installation of mounting kit Mounting the oven to the source Connecting the controller Column loading Software Installation Installing the oven control software Control 11 Using COControl FAQ 13 ...

Page 4: ...ord or plug is damaged liquid has been spilled or objects have fallen into the apparatus the apparatus has been exposed to rain or moisture does not operate normally or has been dropped Use only with accessories specified by the manufacturer or sold with the apparatus Use only power cables rated at least for the maximum input current of the apparatus If the apparatus is used in a manner not specifie...

Page 5: ...arts PRSO V1 BR for Bruker Captive Spray Column Oven consisting of top and bottom part Four conical inserts for column routing Regulator USB to serial converter Software CD USB stick Oven control cable Cable for RS232 and contact closure Power supply cable Power supply 24V wide range input Crossholder with lid Preassembled mounting kit Mounting plate Two rails Slider Mounting screws ...

Page 6: ... two inlets and one outlet for the columns Depending on the column length the front or the lateral column inlet is used and therefore the front or the lateral arm can be used to mount the crossholder It can be used to hold a tee to split away a part of the HPLC flow if needed The conical inserts can be screwed into the bottom part as needed to route the column from the inlet to the outlet Warning D...

Page 7: ... it can be switched between two temparatures with the provided contact closure interface If no serial port is available a USB to serial adapter is provided The front view shows the connector for the connection between controller and oven The rear view shows the connectors for the connection to the power supply the PC and the contact closure interface Fig 02 Controller front view Fig 03 Controller ...

Page 8: ...t to your Captive Spray source There should be some play to align the oven outlet with the inlet union of the source Note Newer Captive Spray sources already have the mounting holes for the column oven If your source does not have the mounting holes please contact Sonation or Bruker Fig 04 Mounting kit installation Fig 05 Bruker Captive Spray Source with mounting holes ...

Page 9: ...24 Volt power supply to the power connector The power supply will be connected to a wall outlet later The wide side of the plug faces the top of the control module do not force it in twisted Connect the controller with the provided RS232 cable to an available com port on your instrument PC The remaining two open leads of this cable are connected to the contact Fig 06 Open oven with Bruker Captive ...

Page 10: ... column either through the front or left side gap depending on the column length Longer columns enter through the left side gap and can be wound around the inserts as often as needed If needed install your T or cross piece into the cross holder It can slide on the left or front rail and can be secured with the Nylon screw in the crossholder Do not use any adhesive tape or other temperature sensiti...

Page 11: ...or message showing NET is missing just install it from the provided CD stick Control Using COControl After starting the software it comes up with the main view Choose the COM port to which the controller is connected and click connect Now you should see the current temperature of the oven in the display in the upper right If no temperature is displayed please ensure that the upper part of the oven...

Page 12: ... displayed To show it up chose Options Graph window from the main menu In the settings window under Options Settings the graph window can be started automatically on startup It is possible to unlock an extended temperature range If your application requires higher or lower temperatures please ask Sonation for an option code But also keep in mind that a higher temperature range is stress for the Pe...

Page 13: ...ensure that your oven is closed properly When the oven is open or even not properly closed the temperature sensor cannot be read Q Is it possible to set up higher temperatures then 50 C or lower temperatures then 30 C A Yes it is possible Please ask Sonation for an option code but remember that the lifetime of the Peltier elements will be lowered because of the additional thermal stress Q My Capti...

Page 14: ...14 PRSO V1 BR for Bruker Captive Spray Users Manual Sonation GmbH Notes ...

Page 15: ...PRSO V1 BR for Bruker Captive Spray Users Manual 15 Sonation GmbH ...

Page 16: ...Sonation GmbH Alte Schulstr 39 88400 Biberach Germany www sonation com Sonation PRSO V1 BR for Bruker Captive Spray ...