32
Radian Motion Sensor
User Manual
Document Ref: UM-8041-C1
Issue: C - Rev 1
© Sonardyne International Limited 2007
Fig 26 Typical MagField Calibration Report screen
The software will inform you if there is insufficient data or if the data is
noisy. If this happens, you must re-start the calibration.
If the calibration has been successful, the report will show how successful it
has been.
You can repeat the calibration if you want to try to improve the result. If you
cannot obtain a better result, then you can re-process the data file that was
saved for the best attempt and then use that. This is an option on the first dia-
log.
12. Close the
Calibration Report
screen.
13. If the results are acceptable, click
Write
to save the results in Radian.