TIME TO PEAK POWER is the time elapsed from stimulus termination to the point of maximum EEG power.
The MAXIMUM SUSTAINED COHERENCE (COH) measure reports the mean value of the 5-second EEG
segment with the highest average coherence recorded during the seizure.
TIME TO PEAK COHERENCE is the time elapsed from stimulus termination to the point of maximum EEG
coherence.
DUKE UNIVERSITY AMPLITUDE MEASURES display the amplitudes of the 3 seizure segments (early ictal,
mid-ictal, and post-ictal) reported by Duke University investigators to correlate with ECT treatment response.
BASELINE RETENTION
Baseline Retention, BLV, found under ENDPOINTS, is the length of time the EEG baseline will be kept in
memory after the treatment. It can be set from 0 - 5 minutes. After this time a new EEG baseline must be
acquired. This feature is useful for re-stimulation without acquiring a new EEG baseline.
TO ENABLE/DISABLE SEIZURE QUALITY MEASURES
FLEXDIAL™ INDEXES: AESI ON/OFF, PSI ON/OFF, MSP
ON/OFF, COH ON/OFF, DUKE ON/OFF
In the “INDEXES” function level, repeatedly
pressing
the
FlexDial
™ will show a sequential flashing status
display (ON/OFF) for ASEI, PSI, MSP, COH, and DUKE measures, in that order.
Rotating
the
FlexDial™
will
flash-display the enable/disable (ON/OFF) options, for each index.
Press
the
FlexDial
™ to select “ON” or
“OFF” for each index and advance to the next. When the last index (DUKE) is enabled/disabled the display
returns to INDEXES function level.
EEG FREQUENCY MEASURES
The Thymatron® System IV provides a continuously updated display of the user’s choice of one of 3 additional
EEG measures: the
95%
Spectral Edge Frequency
(the EEG frequency below which 95% of the total EEG
power is found), the
Median Frequency
(the EEG frequency above and below which 50% of power is found) and
the
Relative Delta Power
(the % EEG power found in the delta bandwidth: 1.5–3.5 Hz).
FLEXDIAL™ EEG FREQ: SP. EDGE, MID FREQ, % DELTA
SP. EDGE
95% Spectral Edge Frequency