I2773-4.
4
en
23
Somat eDAQ
lite
2 Using the eDAQ
lite
The following chapter describes the setup and operation of the eDAQ
lite
including the
LED status indicators, updating firmware, eDAQ
lite
communications, power
considerations, data storage and eDAQ
lite
networking.
2.1
eDAQ
lite
Base System
2.1.1
Status LEDs
There are three LEDs (green, yellow and red) located on the eDAQ
lite
front panel that
are important indicators of eDAQ
lite
status.
Initial State
When powering on, all three LEDs turn on indicating that the main processor is starting
the boot up process. However, the eDAQ
lite
displays different LED states than this on
abnormal conditions as defined by the following table.
NOTE
If all three LEDs stay on for more than about 10 seconds, the boot has failed at a very
early stage or the boot was interrupted in the serial loader (which should not happen
accidentally).
Possible States
After all three LEDs are on, the main processor boot up begins. The LED states that
can exist after the boot up starts are defined as follows.
OO
*55
5VCVWU.'&U
Red
Yellow
Green
Description
4 Hz
Off
Off
Bus pins misaligned.
4 Hz
4 Hz
Off
Blown fuse.
On
On
Off
Power supply failure or voltages not stabilized.
On
On
On
No problems detected, boot up starting.
Red
Yellow
Green
Description
On
On
On
Normal for early in boot up process.
Off
On
On
Normal for later in boot up process.
Off
Off
On
Ready, no test initialized, no error/status flag.
On
Off
On
Ready, no test initialized, error/status flag.
Off
0.5 Hz
On
Ready, test initialized, no error/status flag.
On
0.5 Hz
On
Ready, test initialized, error/status flag.
Off
1 Hz
1 Hz
Ready, waiting for sync, no error/status flag.
On
1 Hz
1 Hz
Ready, waiting for sync, error/status flag.
Off
8 Hz
On
Ready, test running, no error/status flag.
On
8 Hz
On
Ready, test running, error/status flag.
HBM: public
Summary of Contents for eDAQlite
Page 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Page 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Page 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Page 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Page 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Page 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Page 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Page 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Page 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Page 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Page 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...