background image

LCD Replacement Instructions

High Quality Groundwater and Surface Water Monitoring Instrumentation

(Page 1 of 2)

Model 3001 Leveloader Gold/9100 STS Gold

Instructions

1.  Place the Leveloader/Controller device on an anti-static 

table-top mat, and ensure you are wearing an anti-static 

wrist strap.

2.  Use the small screwdriver to remove the battery panel on  

the back of the device, and remove the battery.

3.  Use the Phillips security bit to undo the two security screws 

on the back of the device. 

Tools and Materials Needed

1.  Leveloader Gold/STS Gold Controller LCD (Spare) (#110740)

2.  Screwdriver using Phillips Security Bit

3.  Small Phillips Screwdriver

4.  Tweezers

5.  Anti-static Wrist Strap and Table-top Mat

Leveloader Gold/STS Gold Controller LCD (Spare) (#110740)

4.  Connect a cable (with MD8 end) to the bottom of the 

device, and hold onto the cable for stability, as you pull the 

back housing from the front housing of the device. 

5.  Use the tweezers to carefully pull the tabs on either side of 

the wider ribbon cable connector, towards the bottom of 

the device, to release the ribbon cable from the connector.

6.  Use the tweezers to carefully pull the darker section of the 

narrower ribbon cable connector towards the bottom of the 

device to release the ribbon cable.

Continued overleaf...

Security  

Screws

Battery 

Panel