LCD Replacement Instructions
High Quality Groundwater and Surface Water Monitoring Instrumentation
(Page 1 of 2)
Model 3001 Leveloader Gold/9100 STS Gold
Instructions
1. Place the Leveloader/Controller device on an anti-static
table-top mat, and ensure you are wearing an anti-static
wrist strap.
2. Use the small screwdriver to remove the battery panel on
the back of the device, and remove the battery.
3. Use the Phillips security bit to undo the two security screws
on the back of the device.
Tools and Materials Needed
1. Leveloader Gold/STS Gold Controller LCD (Spare) (#110740)
2. Screwdriver using Phillips Security Bit
3. Small Phillips Screwdriver
4. Tweezers
5. Anti-static Wrist Strap and Table-top Mat
Leveloader Gold/STS Gold Controller LCD (Spare) (#110740)
4. Connect a cable (with MD8 end) to the bottom of the
device, and hold onto the cable for stability, as you pull the
back housing from the front housing of the device.
5. Use the tweezers to carefully pull the tabs on either side of
the wider ribbon cable connector, towards the bottom of
the device, to release the ribbon cable from the connector.
6. Use the tweezers to carefully pull the darker section of the
narrower ribbon cable connector towards the bottom of the
device to release the ribbon cable.
Continued overleaf...
Security
Screws
Battery
Panel