Solinst 201 Instructions Download Page 1

Water Level Temperature Meter Instructions

High Quality Groundwater and Surface Water Monitoring Instrumentation

(Page 1 of 2)

Model 201

Operating Principles

The Solinst Model 201 Water Level Temperature (WLT) Meter 

measures water level and temperature. Temperature readings are 

displayed on the LCD screen and water level is read from the tape as 

with a conventional Water Level Meter. When the probe is immersed in 

a conductive fluid, a circuit is completed and the water level is indicated 

by a tone and light. 

Equipment Check

Upon receipt of your Solinst Model 201 Water Level Temperature 

Meter, and always before heading out to the field, the following checks 

are recommended:

1.  Press the button to turn the meter on. The display should briefly 

show the firmware version of the meter (e.g. WLTM 1.0), then the 

air temperature (e.g. 21.12 C). If the battery is low, a ‘LOW BATT’ 

warning appears and the 9 volt alkaline battery should be replaced. 

If “No Comm” appears, check the probe connection (call Solinst if 

message persists). If the display is blank, install a new 9V battery.

2.  Ensure the probe tip is clean.
3.  Submerse the probe in tap water. This completes the circuit and 

activates the buzzer and light.

Tape Guide

Brake

Probe Holder

Front View

Back View

 

LCD Display

Indicator 

Light

Battery 

 Drawer

Handle

Sonalert 

(Buzzer)

Button

1.  Turn the meter on and lower the probe into water. A tone and light 

indicate that water has been reached and the depth can be read 

off the tape and recorded. Lower and raise the probe slowly a few 

times to verify the depth.

2.  Once in water, the screen displays temperature of the water at the 

zero point. Lower the probe to the desired depth. Record the depth 

and the associated temperature reading when stabilized. 

Taking Water Level & Temperature Measurements

Notes:

1. The zero measurement point of the Water Level Temperature Meter 

is located near the tip of the probe.

2. If the display indicates ‘LOW BATT’ there is still some life left in the 

battery, but it is recommended that you change the batteries as soon 

as possible.

3.  To turn the buzzer off while you are obtaining a temperature 

profile, press the button 2 times quickly. “WL-off” will display 

above the temperature reading on the LCD. To turn the buzzer 

back on, press the button 2 times quickly.  

4.  To conserve battery power, the display has an auto off after  

8 minutes of use. If the display is blank when you wish to take 

a measurement, press the button to display temperature reading.

5.   Repeat at each desired depth.
6.   After each use clean the sensor pin with a soft cloth and rinse with 

deionized water.

7.  To turn the Water Level Temperature Meter off, press and hold the 

button for about 2 seconds. 

Tape Guide Instructions

1.   Fit the Tape Guide over the top of the well, small end in.
2.   Insert the leg of the Water Level Temperature Meter into the hole 

on the Tape Guide and rest the Water Level Temperature Meter on 

the side of the well casing (small reels only, see diagram).

3.   Take all measurements at the ‘V’ notch on the Tape Guide, and 

adjust readings according to the offset stamped on the Tape Guide 

(i.e. subtract 6 cm or 2/10 ft).

4.   When finished, store the Tape Guide by clipping it onto the support 

bracket on the back of the Water Level Temperature Meter.

Solinst Canada Ltd.

www

.solinst.com

Tel:

 +1(905) 873-2255

(800) 661-2023

Tape Guide:

 prevents tape 

from scratching the top edge 

of well casing 

Water Level  

Temperature Meter: 

anchored to the well casing with the 

tape guide for stable operation. Small 

reels only. Medium and large reels 

should be placed on the ground.

High quality PVDF flat tape: 

hangs straight in the well for 

accurate interface  

measurements

Note:

 

Temperature readings are most accurate when the entire probe 

is submersed. 

Zero Point