Solaxx SafeDip Operating Instructions Download Page 1

Digital Chemistry Reader

Meter for Swimming Pools and Spas:

· Free Chlorine
· pH
· Salinity
· TDS
· ORP 
· Temperature

WARNING! 

This  package  contains  pH  calibration  chemicals  that  may  be 
harmful if misused.   Please read instructions carefully.   Not to be 
used by children except under adult supervision.

INTRODUCTION

S A F E - D I P ™   i s   a   s w i m m i n g   p o o l   w a t e r   t e s t e r   o f  
Cl/pH/Salt/TDS/ORP (Free Chlorine/pH/Salinity/TDS/ ORP). The 
SAFE-DIP™  is  an  easy  to  use,  accurate,  and  economical 
measurement device that offers free chlorine level indication and 
direct reading of pH, salinity, TDS and ORP with one device.  It is a 
device  that  covers  almost  all  of  the  most  important  pool  water 
chemistry in order to maintain the water at its optimal condition for 
bathers' safety and comfort as well.

Introduction & Safety Information

Specifications / Features

SAFE-DIP™ Meter Includes:

pH 7.0 Buffer Packet (for calibration)

PARTS & ACCESSORIES

pH Replacement Electrode
pH 7.0 Buffer Packet

Ÿ

  Operating range: 5~50°C, 0~12pH, 0 ~ 999mV (ORP), 

1~80mS (Conductivity)

Ÿ

  Replaceable pH sensor. Resolution: 0.1 pH

Ÿ

  1 point pH calibration using 7.00pH standard buffer

Ÿ

  Automatic temperature compensation for pH and conductivity

Ÿ

  Temperature sensor. Resolution: 1°C; Accuracy: ±5% (~±1°C), 

Selectable °C/°F system

Ÿ

  ORP sensor. Resolution: 1mV; Accuracy: ±20mV 

(500~800mV) 

Ÿ

    TDS  and  SALT.  Resolution:  100ppm;  Accuracy:  ±20% 

(200~5000ppm)

Important:

Please  refer  to 

www.safedip.com

  for  a  more  detailed  operator 

instruction. The detailed instruction includes digital images in full 
color as well as an online calculator which tells you exactly what to 
add to your water based on the data retrieved from the Safedip 
digital reader.

Go to 

www.safedip.com

 for more detailed instructions. 

Front Panel Description 

1.   Battery compartment cap
2.   LCD Display
3.   ON/OFF button 
4.   START button – start a new measurement
5.   CAL button – calibration of pH probe

o

o

6.    C/ F button – toggle between Celsius and 
       Fahrenheit
7.   Sample Cup and 
      Electrode Compartment (inside)
8.   Probe Electrode
9.   Sleeve holding area (when meter is in use)
10. Probe Sleeve

Cl

pH

Salt

TDS

ORP:

:

:

:

: Good Low High

ppm

ppm

mV

88

.

8800

8800

888

88

C

F

1

2

3

4

5

6

7

8
9

10

11

12

8

1

2

3

5

4

6

7

9

SAFE-DIP™ Display  

1.   Main display
2.   Cl (free chlorine) indication Display Row 
3.   pH measurement Display Row
4.   Salt (salinity) measurement Display Row 
5.   TDS measurement Display Row 
6.   ORP measurement Display Row
7.   Hourglass: measurement in progress indicator
8.   Battery indicator
9.   Check mark: measurement done indicator
10. Disk icon: readings from memory
11. Tools icon: error warning
12. Temperature display

Meter Description

Go to 

www.safedip.com

 for more detailed instructions. 

Getting Started

Setting up the meter

1.  Place  batteries  (included)  into  the  device.  Refer  to  Battery 
Replacement section of this manual.
2. Remove the sample cup in order to access the probes.
3. Remove the sleeve which protects the probes.
4. Calibrate the probe prior to first use and every month by referring to 
the Calibration section of this manual.

Measurement

1.  Fill the bottom 1/3 of the sample cup with the water sample.  Make 
sure  to  collect  the  water  at  least  1.5”  from  the  surface.  The  other 
option is to submerse the probes directly into the water sample. 
2.  Press the   button. 
The  most  recent  measurement  will  always  appear  along  with  the 

 icon.

3.   Remove the sleeve and attach it to holding area. Immerse the 
Safedip  probe  in  the  water  sample.    Make  sure  the  probes  are 
completely submersed. 
4.  Press the START button to begin a new measurement.
5.  During the new measurement, the   will blink before all stabilized 
parameters are achieved. 
6.  When measurement is finished, a beep will be heard and newest 
data will be displayed. The data will be saved in the memory until the 
next  measurement.  Only  the  latest  measurement  is  stored  and  all 
prior measurements are discarded. 
7.  In order to view the latest readings at anytime, press     .
8.  If a new measurement is desired, repeat steps 3-6.
9.   When finished using the meter, press the               once to turn the 
device off. If you forget to do so, the meter will turn off automatically 
after 1 minute.
10. Always rinse the probes with fresh tap water after use. Replace 
the sleeve in order to keep the probes wet. Failure to keep probes wet 
will void the warranty.

   

     

 

 

___ 

Basic Operation

1)  Power ON/OFF:

The SAFE-DIP™ uses two AAA batteries. If the batteries are weak, 
the

 icon indicator will appear on the display. Press the ON/OFF 

button to turn the SAFE-DIP™ on or off. 

The auto power off feature 

will turn off the SAFE-DIP™ automatically after 1 minute of non-
use if no buttons are pressed.

____

o

o

2)  Toggle  C/ F:

To change the displayed temperature units between °C or °F:

o

o

With the SAFE-DIP™ ON, press  C/ F button .

3)  New Measurement:

When  the  meter  is  powered  on,  the  LCD  display  will  display  the 
readings from the very last measurement.   The         icon will also be 
displayed.  Press START button to start a new measurement.  During 
the new measurement, the     icon will blink before all stabilized data 
are achieved.  (Please see detailed explanation in the Measurement 
Procedure section.)

Go to 

www.safedip.com

 for more detailed instructions. 