64
In weighing mode, place the item to be weighed and store the weight value in the
picking memory by pressing the Plus button.
The display returns to zero.
Place the next item on the weighing platform and store the weight in the
order-picking memory by pressing the Plus button.
Repeat this process as often as you like.
A full commissioning memory can be recognised by the total symbol (Σ) in the
display.
Recalling the total value
First press the INFO key and then the Plus key.
There is a rolling display: Gross weight, tare weight, net weight, piece total,
number of picks and flashing symbol „Σ“. Cancel or exit with the On/Off key.
Clear the picking memory
Press the Clear button. With other UCAL setting, by CL and Plus key.
Picking with tare
See point 7.3.
An active tare weight can be recognised by the „NET“ in the display.
Auto-picking
(see 470.702.126 User mode)
In weighing mode, place the goods to be weighed on the scale and store the first
weight value in the picking memory by pressing the Plus key.
Remove the weighing product from the platform and place the next weighing
product on it. The weight value is automatically stored in the order-picking
memory. Repeat this process as required.
9. Application programmes
9.6 Order picking
9.7 Neutral measurement
In setting mode, „Neutral measurement“ must be selected for the assignment of
the function button! For setting the function button see chapter 8.1 or user
mode 3820 (see separate description 470.702.126) UCAL 4 pos. 02 = value 8.
The factor for entering the specific weight is the quantity of the substance to be
measured in relation to the calibrated unit (1 kg/1 g/1 lb).
Example: 1 kg paint = 0.823 litre (input UCAL1, pos. 27 = 0.823).
Entering the factor for the conversion in the setting mode
Press SEND- and then together with the On/Off button for 5 seconds until
appears.