57
OPTION
AL
FUNC
TIONS
8
8.7 Tenfold resolution x10
The tenfold higher resolution appears by pressing the SEND and then the
and then the zeroing button at the same time.
On scales calibrated for legal metrology, the higher resolution appears flashing
for 5 seconds.
For non-verifiable calibrated scales, the higher resolution appears until the
SEND and zeroing buttons are pressed again, until the SEND and zeroing
buttons are pressed again as when the scale was called up.
8.8 Convert weight units kg /lb
The possibility of weight unit changeover must be enabled in calibration mode
(ECAL 1 pos. 05 value = 0)!
A g - kg - lb changeover is only possible with the non-verifiable version!
Conversion of second unit (example to lb) in UCAL 01 pos. 04 value = 2:
Press the SEND button and then the On/Off button for 5 seconds until appears.
Accept with the SEND button.
Then clock up to position 04 with the tare button and confirm with the SEND
button. confirm with the SEND button.
Set to value range 2 with the tare button and confirm again with the SEND
button.
Position 05 is shown in the display.
Press the zeroing button; this is shown in the display.
Quit and save the setting with the SEND button and then briefly together with the
zeroing button, display returns to weighing mode.
Switching over:
Press the function button to change over the unit.
For changing over the function button, see chapter 8.1.
8. Optional functions