55
OPTION
AL
FUNC
TIONS
8
8. Optional functions
8.5 Alibi memory (option for legal-for-trade data transmission)
The Alibi memory is activated in the setting mode (see separate description
470.702.126 User mode). When the alibi memory is activated, the corresponding
symbol appears. The symbol flashes as soon as the entered limit value for full
message is reached. 1.
1. Transfer to Alibi memory
By requesting the ADV data record „IT in Alibis“ or by pressing the SEND button,
the weight is taken over into the alibi memory. However, the printout or IT data set
must be configured accordingly (see separate description 470.508.108 Data
interface). 2.
2. Query alibi memory
The alibi memory is queried in the setting mode.
Press and hold the SEND button and then the On/Off button for approx. 5 seconds
until UCAL1 appears.
Toggle in the menu with the tare button until UCAL5 appears and confirm with the
SEND button.
Then advance with the tare button to setting step 03 and confirm again with the
SEND button.
The number of the last data set saved appears.
You can move through the alibi memory using the tare button and the function
button and call up each data record using the SEND button.
The selected data record is displayed in a rolling sequence as follows:
Indication in the
display
Meaning
12
Sequence number of the alibi entry
t2890_01
Scale type
Serial number
of the measuring
point
y 22
Year
1234
Consecutive number
A 12.34
kg
Gross or net with alibi
A 2,98
T
kg
T
Tare with alibi
Scroll further using the tare and function buttons.