71
Scope Multimeter Operations
Getting Started
8.2.1 Capabilities
The following tables detail the software and hardware capabilities.
Table 8-1 Scope
Function
Range
Accuracy/Comments
Signal Measurement
Ch. 1—yellow banana jack
Ch. 2—green banana jack
Ch. 3—blue banana jack
Ch. 4—red banana jack
Each channel input is referenced to
common ground input (GND—black banana
jack)
Sample Rate
Single trace mode = 6 MSPS
(Ch. 1 only)
Dual trace mode = 3 MSPS
(Ch. 1 & 2 only)
Four trace mode= 1.5 MSPS
(Ch. 1–4)
Continuous sampling,
MSPS = mega samples per second
Bandwidth
DC–3 MHz
3 db point @ 3 MHz
Input Impedance
10 MOhm
@ DC
4 kOhm @ 3 MHz
Channel 1–4
VDC (Full Scale)
100mV–400V
Do not measure greater than 75VDC
VAC (Full Scale)
100mV–400V
Do not measure greater than 50VDC (rms)
Table 8-2 Graphing Multimeter
Function
Range
Accuracy/Comments
Signal Measurement
Ch. 1—yellow banana jack
Ch. 2—green banana jack
Each channel input is referenced to
common ground input (GND—black banana
jack)
Sample Rate
1.5 MSPS
Continuous sampling,
MSPS = mega samples per second
Bandwidth
DC–3 MHz
3 db point @ 3 MHz
Input Impedance
10 MOhm
@ DC
4 kOhm @ 3 MHz
Channel 1–4
VDC (Full Scale)
75VDC
Do not measure greater than 75VDC
VAC (Full Scale)
50VAC
Do not measure greater than 50VDC (rms)
Low Amp Probe
20A scale (100mV/Amp)
40A scale (10mV/Amp)
Connect Low amp Probe (+) to Ch. 1
(yellow banana jack) and (–) to GND input
(black banana jack).
See NOTE below.
Table 8-3 Digital Multimeter
Function
Range
Accuracy/Comments
Signal Measurement Ch. 1—yellow banana jack
Channel 1 input is referenced to common
ground (GND—black banana jack)
VDC (Full Scale)
75VDC
Do not measure greater than 75VDC
VAC (Full Scale)
50VAC
Do not measure greater than 50VDC (rms)
Signal Measurement
Input Impedance
10 MOhm
Ohm Measurement
Diode Test
Continuity Test
Ch. 3—blue banana jack
Ch. 4—red banana jack
Each channel input is referenced to
common ground (GND—black banana jack)
Ohms
400 Ohm–40 MOhm
Fixed scales or Auto Ranging
Glitch capture
Approximately 50 uS
Diode Test
2 V Scale