Safety Notes

 

 

• 

The Volaro lift is designed for patient transfer only.  Using the lift for transport can create an unsafe   

patient handling situation.

 

 

• 

Lift legs must be fully extended into the wide position when lifting a patient or resident.

 

 

• 

For Emergency Stop, pull power pac from power box.

 

 

• 

Make sure both loops from the sling are properly 

nested

” 

in the bottom of the hooks before lifting or 

transferring a patient or resident.  Also make sure that both retainer springs are functioning correctly.

 

 

• 

Use 

only 

Volaro slings and accessories designed for use with the Volaro stand models.

 

 

• 

Do not

 push or pull on the beam or resident. This will jeopardize the lift

s stability.

 

 

• 

Do not

 use any other power pac charger. use only the power pac charger supplied with the lift.

 

 

• 

Do not

 exceed the weight limit listed on the lift.

 

 

• 

Do not

 use a sling that shows wear, is torn, bleached out or has loose threads.  Inspect sling before each 

lift.

 

 

 

 

PARTS & SERVICE

 

 

SMT Health Systems

 

 

Phone: (218) 847

-

3386  Toll Free: (800) 725

-

7761 

 

Toll Free Fax: (888) 385

-

2345

 

Email: [email protected]

 

 

1380 Legion Road, Detroit Lakes, MN 56501

 

2

 

Safe Working Loads

 

 

Volaro slings have been developed, tested and manufactured to have a safe working load of 1000 lbs.  Due 

to the variety of resident shapes and dimensions, the appropriate size should be selected to accommodate 

specific patients.   Care should be taken to ensure that the mechanical lift selected has the capacity to safely 

lift the resident.

 

 

The Volaro Sling Sizing Chart (

page ?

) is designed only to suggest which resident will safely fit into the 

sling based on their size and weight.

 

 

NOTE:

 Lifts/Stands are designed to 

transfer

” 

a resident, not 

transport

” 

them.  

 

Summary of Contents for Volaro Series 4 Stand

Page 1: ...PA600 Operator s Manual SMT Health Systems 1380 Legion Road Detroit Lakes MN 56501 800 725 7761 www patientlift com Volaro Series 4 Stand...

Page 2: ...er supplied with the lift Do not exceed the weight limit listed on the lift Do not use a sling that shows wear is torn bleached out or has loose threads Inspect sling before each lift PARTS SERVICE SM...

Page 3: ..._____ _____________________________Date_____________ Assessment Care Planned by_________________________Date_____________ Can individual safely perform or 50 of lift transfer with only standby assista...

Page 4: ...Diameter Screw Vertical Sliding Adjustable Shin Rest adjust ment range of 9 25 inches Horizontal Sliding Adjustable Shin Rest adjustment range of 4 inches 90 Degree Self Pivoting Shin Rest Steel Hand...

Page 5: ......

Page 6: ...ashing instructions are as follows Brush with warm soapy water and non chlorine disinfectant Rinse Drip dry only The sling rigidity makes it more difficult to machine wash Care should be taken if atte...

Page 7: ...7...

Page 8: ...ar down as you can Position the wings of the sling under the arms and make sure they are centered Secure buckle as low as possible around the lower abdomen Fasten the buckle and pull snug Tighten by h...

Page 9: ...eet off the foot board Be sure to use proper body mechanics by assisting the person to a lying position Transferring from Bed to Chair When transferring from bed to wheelchair it is required that the...

Page 10: ...ce to operate the lift This allows the operator to be next to the person being transferred and to operate the lift at the same time It also aids as a back up source to operate the lift Figure C Emerge...

Page 11: ...cause it to overcharge If the power pac goes to GREEN in a short period it may mean you have a weak cell that will not take a full charge Plug power pac into charger every night Do not remove until ch...

Page 12: ...VOLARO STAND PA600 PA600S PA600 EXPLODED VIEW...

Page 13: ...PA600 PARTS LIST...

Page 14: ...age them Circuit Board Wiring Diagram 1 Positive up to Switch Small Green 2 Positive Power Large Red 3 Negative Power Large White 4 Positive to Switch 2 Small Red 5 Safety Switch 2 Small White 6 Up Th...

Page 15: ...ac 1 Power pac runs down quickly 2 Power pac light stays on 1 Power pac is worn out or not being charged when red charge light comes on lift 2 Power pac is worn out or circuit board is bad 1 Make sure...

Page 16: ...ds 4 Get supplies ready sling wheelchair blanket or lap robe 5 Get help from experienced team member for this procedure 6 Have assistant stand on opposite side of bed to assist with transfer 7 Provide...

Page 17: ...member to adjust the safety strap as the resident comes up to keep it snug at all times 20 Lift resident until buttocks have cleared the bed Give assurance to the resident 21 Pull the lift clear form...

Page 18: ...Check the movement of the lift remove hair from the casters if needed Check leg covers if cracked replace with new ones Check all vinyl covers if worn replace with new ones Check all external fittings...

Page 19: ...ut 8 Actuator Noise Up Down Replace bearings after 5 years Grease should be greased at least semi annually 9 Power Box Plastic Guides in Place Terminals Remote Jack Holder or Lens 10 Thumb Switches Le...

Page 20: ...this form for Each Lift Date Checked By Repairs Needed Completed SMT Health Systems www patientlift com sales smthealthsystems com 800 725 7761 Lift Inspection must be done Monthly sling included Mod...

Page 21: ...Motor Actuator Any Other Moving Parts Limited Warranty on all parts only under conditions of normal and intended use Main Structure Main Integrity and Welds of Frame Base Legs 2YEAR Limited Slings Ma...