Chapter 3: Setting Up the Monitor
3-4
Operation Manual SurgiVet
®
V3404 Plus
Adjusting Waveform Sweep Time, Size or Scales
The sweep time, height, and scales for waveforms can be adjusted as follows:
The ECG waveform sweep speed can be adjusted in the
SETUP/VOLUME
menu. The waveform scale can be
adjusted in the
ECG
menu.
The Plethysmogram sweep speed can be adjusted in the
SETUP/VOLUME
menu. The waveform is scaled
automatically to fit the display area.
Each parameter’s trend display scales can be adjusted separately through the
TRENDS
,
SCALES
menu.
Adjusting the Line Filter Parameter
The line filter needs to be adjusted based on the frequency of the line voltage being supplied to the monitor.
For 50Hz line voltage frequency, select the 50Hz line filter option. For the 60Hz line voltage frequency, select the
60Hz line filter option. The line filter frequency can be adjusted as follows:
Press the Wave/Trend key
a
and the Menu/Enter key
b
simultaneously to enter the service mode.
Use the ARROW (
on
) keys to select the line filter option.
Select the frequency that matches the input line voltage frequency.
1.
2.
3.
1.
2.
3.