S E C T I O N 1 0
•
M a i n t e n a n c e
HOTLINE
®
Blood and Fluid Warmer | Operator’s Manual
31
6
Remove the fill-port plug (
c
), fill the reservoir with the hydrogen
peroxide solution, and replace the fill-port plug.
7
Remove the reflux plug from the socket if required, and insert a
HOTLINE
Fluid Warming Set (
d
) (L-70, L-70NI, L-80) into the
socket.
8
Turn the HOTLINE
Warmer ON, and let the recirculating
solution circulate for a 30-minute disinfection period.
9
Turn the HOTLINE
Warmer OFF and unplug the power cord.
10
Empty the reservoir.
11
Remove the HOTLINE
Fluid Warming Set and discard
according to established hospital procedures.
These suggested instructions are designed to be used in conjunction
with established hospital procedures.
Add Recirculating Solution
CAUTIONS
• Do not fill the HOTLINE
Warmer reservoir with a HOTLINE
Fluid Warming Set or a TEMP CHECK in place. Failure to
remove the HOTLINE
Fluid Warming Set before the fill
procedure may result in an air lock in the HOTLINE
Warmer.
1
Prepare the recirculating solution.
2
Remove the fill-port plug.
3
Fill
the reservoir with 1.4 liters of recirculating solution.
4
Replace the fill-port plug.
Maintenance Performed Every 30 Days
Disinfect the Reservoir and Change Recirculating Solution for
Distilled Water and 35% Isopropyl Alcohol Solution
Refer to
Disinfect the Reservoir and Change the Recirculating Solution
procedure in this section.
o
o
C
C
HOTLINE
Fluid Warmer
d