Smiths Medical Level 1 H-1200 Step-By-Step Manual Download Page 1

6

 

Insert the Patient Line in the 

Clamp Slot and push it back  

into the slot. 

15

 

Make patient connection.

9

 

Close pinch clamps above the 

Heat Exchanger and below the  

IV Fluid bag spikes.

2

 

Snap the Heat Exchanger  

  into Guide.

11

 

Hang spiked IV Fluid bags in 

Pressure Chambers.

 Choose the hook that allows the bag port  

   to hang freely in the indented slot at the  
   bottom of the chamber door.

Close the doors and secure the  

latches.

Push the bottom end of the Heat  

  Exchanger firmly into #1 block.

3

 

Slide #2 block down firmly.

10

 

Remove air from IV Fluid  

bags:

 Using aeseptic technique, pierce bag  

   port membrane with disposable set  
   spike.

 Withdraw spike, squeeze out ALL air,  

   and re-spike bag.

4

 

Move the pinch clamp connector  

on the Patient Line of the Gas Vent/

Filter Assembly next to the Luer 

connector. Close the pinch clamp.

7

 

Push door up to engage the top 

hinge and then push door in to 

engage the metal tab to close.

12

 

Prime through the Gas Vent/

Filter Assembly:

 Open pinch clamp/s below IV Fluid bag 

spike, squeeze drip chamber and prime 
the line through the Gas Vent/Filter  
Assembly.

 Open remaining pinch clamp/s below IV 

Fluid bag spike/s to prime additional line/s.  

 When lines are primed, close pinch  

clamps, leaving one open for infusion.

When IV Fluid bag empties:

 1.  Move the Pressure Chamber lever to 

the 

 position.

 2.  Close the pinch clamp below empty  

IV Fluid bag.

 3.  Move lever on the other Pressure 

Chamber to 

+

 position and open 

pinch clamp to begin infusion.

 4.  Open door of de-pressurized  

Chamber and replace empty IV Fluid 
bag with new bag with ALL air  
removed (Steps 10 & 11).

 5.  Alternate between Pressure  

Chambers.

5

 

Open #3 Clamp Slot door. Press  

down on the metal tab and pull door 

down and away from Clamp Slot.

13

 

Open pinch clamp below Gas 

Vent/Filter Assembly to prime the  

patient line and remove all air, 

and then close the roller clamp.

8

 

Press Gas Vent/Filter Assembly 

into #4 block.

14

 

Press the green Power ON  

button.

16

 

Open roller clamp on the  

patient line to begin infusion.

IMPORTANT:  Please refer to the Operator’s Manual supplied with the product for detailed instructions, contraindications, warnings, and cautions.  

Refer to the IFU for DI-60HL Administration Set for specific priming instructions.

17

 

Move Pressure Chamber  

lever to the 

+

 position.

Level 1

®

 H-1200 Fast Flow Fluid Warmer 

DI-Series Disposable IV Administration Set 

(DI-70/DI-100/DI-300/DI-60HL)

Step-by-Step Guide